47
•
Chapter 7: Maintenance
65-2646XL-05-04 M2A Transmitter Operator’s Manual
7. Connect the detector leads to the terminal block the same way the old detector was wired (see
8. Screw the calibration adapter/splash guard onto the new detector housing cap.
9. Reinstall the detector junction box cover.
10. Turn on or plug in power to the M2A.
CAUTION:
Allow the replacement detector to warm up for 5 minutes
before you
continue with the next step.
11. Calibrate the replacement detector as described in “Calibration” on page 41.