65-2640RK-03 M2A Transmitter Operator’s Manual
39
10. Turn on or plug in power to the M2A.
NOTE:
Allow the replacement detector to warm up for 15 minutes
before you continue with the next step.
11. Calibrate the replacement detector as described in “Calibration” on page 36.