background image

65-2451RK Hydrogen Transmitter • 3

Amplifier

The amplifier converts the electrical output from the detector to a 4 to 20 mA signal that 
corresponds to the detection range and transmits the signal to a gas monitoring controller. 
A foam gasket that orients the amplifier and keeps it from rotating is installed on the 
bottom of the amplifier. A label on the amplifier indicates the detector drive current. This 
drive current is factory set and is dictated by the combustible gas to which the detector is 
calibrated. Consult RKI Instruments, Inc. if you plan to change the calibration gas for the 
transmitter, for example from a hydrogen calibration to a methane calibration. The 
amplifier includes the detector terminal strip, controller terminal strip, span pot, zero pot, 
and test points (see Figure 1).

Controller Terminal Strip

The controller terminal strip is a three position plug-in style terminal strip located near the 
top of the amplifier. Use the controller terminal strip to wire the transmitter to a controller.

Detector Terminal Strip

The detector terminal strip is a four position plug-in style terminal strip located below the 
controller terminal strip. Use the detector terminal strip to connect the detector to the 
amplifier.

NOTE:

The detector is factory-wired to the detector terminal strip. See “Wiring the 
Hydrogen Transmitter to a Controller” on page 5 for all wiring procedures 
related to the transmitter.

Zero Pot

The zero pot is located in the upper left corner of the amplifier (see Figure 1). Use a small 
flat blade screwdriver to turn the zero pot’s adjustment screw and adjust the amplifier’s 
zero (fresh air) output during the start-up and calibration procedures. Turn the adjust-
ment screw clockwise to increase the zero output and counterclockwise to decrease the 
zero output.

Span Pot

The span pot is located to the right of the zero pot (see Figure 1). Use a small flat blade 
screwdriver to turn the span pot’s adjustment screw and adjust the amplifier’s gas 
response output during the calibration procedure. Turn the adjustment screw clockwise to 
increase the zero output and counterclockwise to decrease the zero output.

CAUTION:

The amplifier includes two additional potentiometers. They are factory-set. Do not 
adjust them.

Test Points

The test points are on the left side of the amplifier (see Figure 1). The test points produce a 
100 mV to 500 mV output that corresponds to the transmitter’s 4 to 20 mA output. Use the 
test points and a voltmeter to measure the amplifier’s output during the start-up and 
calibration procedures. The black test point in the lower left corner is the negative (-) test 
point and the red test point below the zero pot on is the positive (+) test point.

Junction Box

Use the junction box to install the transmitter at a mounting site that is remote from the 
controller. The junction box protects the amplifier and wiring connections made to the 
amplifier. Use the top 3/4 inch conduit hub to connect wiring from the amplifier to the 

www.

GlobalTestSupply

.com

Find Quality Products Online at:

[email protected]

Содержание 65-2451RK

Страница 1: ...65 2451RK Hydrogen Transmitter Operator s Manual Part Number 71 0133RK Revision C Released 5 17 11 www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply com...

Страница 2: ...itter is essential for proper operation and correct readings Please calibrate and maintain this transmitter regularly Frequency of calibration depends upon the type of use you have and the sensor type...

Страница 3: ...n costs or contingent expenses incurred without our prior approval THIS WARRANTY IS EXPRESSLY IN LIEU OF ANY AND ALL OTHER WARRANTIES AND REPRESENTATIONS EXPRESSED OR IMPLIED AND ALL OTHER OBLIGATIONS...

Страница 4: ...up 7 Introducing Incoming Power 7 Setting the Zero Signal 7 Maintenance 8 Preventive Maintenance 8 Troubleshooting 9 Replacing Components of the Hydrogen Transmitter 10 Calibration Frequency 12 Calibr...

Страница 5: ...ing the 65 2451RK you must follow the instructions and warnings in this manual to assure proper and safe operation of the 65 2451RK and to minimize the risk of personal injury Be sure to maintain and...

Страница 6: ...only allows hydrogen to diffuse into the detector The 1 2 inch NPT mounting threads at the top of the detector allow you to mount it into the bottom conduit hub of the junction box A rainshield screw...

Страница 7: ...e zero pot is located in the upper left corner of the amplifier see Figure 1 Use a small flat blade screwdriver to turn the zero pot s adjustment screw and adjust the amplifier s zero fresh air output...

Страница 8: ...troller Mounting the Hydrogen Transmitter 1 Select a mounting site that is representative of the monitoring environment Consider the following when you select the mounting site Select a site where the...

Страница 9: ...tion box 7 Pull out the controller terminal strip and connect the three wires to the terminal strip as follows see Figure 3 Connect the positive wire to the PWR SIG terminal Connect the signal wire to...

Страница 10: ...ed connect the cable s drain wire to an available chassis earth ground at the controller Controller or Recording Device 24 VDC Amplifier Black Red Cable Shield DC Ground 4 20 mA In S PWR SI G S R W G...

Страница 11: ...ing operation CAUTION If you suspect the presence of hydrogen in the monitoring environment use the calibration kit and the zero air calibration cylinder to introduce fresh air to the detector and ver...

Страница 12: ...e performing this test NOTE The following procedure assumes the use of a calibration kit which includes a calibration gas cylinder a 0 5 LPM fixed flow regulator with an on off knob a calibration cup...

Страница 13: ...he regulator 6 Unscrew the regulator from the calibration cylinder 7 Unscrew the calibration cup from the detector NOTE For convenience leave the regulator and calibration cup connected by the sample...

Страница 14: ...te the transmitter 3 If the fail condition continues replace the detector 4 If the fail condition continues contact RKI for further instruction Slow or No Response Difficult or Unable to Calibrate Tra...

Страница 15: ...the amplifier to the junction box 6 Remove the old amplifier 7 Install the amplifier into the junction box with the screw lock washer and flat washer you removed in Step 5 above A foam gasket that ori...

Страница 16: ...n dictates otherwise RKI Instruments Inc recommends a calibration frequency of every 3 months If an application is not very demanding for example detection in a clean temperature controlled environmen...

Страница 17: ...gative lead into the black test point 5 Use the following formula to determine the correct test points output for the calibrating sample Output mV calibrating sample fullscale X 400 100 For example wi...

Страница 18: ...e reading to match the correct response reading 4 Turn the regulator knob clockwise to close the regulator 5 Unscrew the regulator from the calibration cylinder Returning to Normal Operation 1 Remove...

Страница 19: ...es detector junction box and amplifier 71 0133RK 65 2451RK Hydrogen Transmitter Operator s Manual this document 81 0002RK 01 Steel calibration cylinder 50 LEL hydrogen in air 34 liter 81 0076RK 01 Ste...

Отзывы: