background image

65-2434RK CO Transmitter • 9

Symptoms

The transmitter responds slowly or does not respond during the monthly response 
test.

Unable to accurately set the zero or response reading during the calibration 
procedure.

The transmitter requires frequent calibration.

NOTE:

Under “normal” circumstances, the transmitter requires calibration once every 
three months. Some applications may require a more frequent calibration 
schedule.

Probable causes

The calibration cylinder is low, out-dated, or defective.

The transmitter is malfunctioning.

Recommended action

1.

Verify that the calibration cylinder contains an adequate supply of a fresh test sample.

2.

If the calibration/response difficulties continue, replace the CO sensor as described 
later in this section.

3.

If the calibration/response difficulties continue, contact RKI Instruments, Inc. for 
further instruction.

Replacing Components of the CO Transmitter

This section includes procedures to replace the CO sensor, amplifier PCB, and charcoal 
filter. 

Replacing the Sensor

1.

Turn off the controller.

2.

Turn off or unplug incoming power to the controller.

3.

Remove the enclosure’s cover by unscrewing the four cover screws, one at each 
corner.

4.

Unplug the CO sensor from the amplifier PCB.

5.

Remove the rubber boot and charcoal filter from the old sensor and install onto the 
replacement sensor.

6.

Carefully plug the replacement sensor into the socket pattern on the amplifier PCB.

NOTE:

Match the sensor’s male pins with the four female sockets as you plug the sensor 
into the sockets.

7.

Secure the cover to the enclosure.

8.

Turn on or plug in incoming power to the controller.

9.

Turn on the controller.

CAUTION:

Allow the replacement sensor to warm up for 5 minutes before you continue with the 
next step.

10. Calibrate the replacement sensor as described in the Calibration section of this 

manual.

www.

.com

[email protected]

1.800.561.8187

Содержание 65-2434RK

Страница 1: ...65 2434RK Carbon Monoxide Transmitter Operator s Manual Part Number 71 0061RK Revision D Released 2 16 11 www com information itm com 1 800 561 8187...

Страница 2: ...nce of the detector is essential for proper operation and correct readings Please calibrate and maintain this detector regularly Frequency of calibration depends upon the type of use you have and the...

Страница 3: ...costs transportation costs or contingent expenses incurred without our prior approval THIS WARRANTY IS EXPRESSLY IN LIEU OF ANY AND ALL OTHER WARRANTIES AND REPRESENTATIONS EXPRESSED OR IMPLIED AND AL...

Страница 4: ...tter to a Controller 5 Startup 6 Introducing Incoming Power 6 Setting the Zero Signal 6 Maintenance 7 Preventive Maintenance 7 Troubleshooting 8 Replacing Components of the CO Transmitter 9 Calibratio...

Страница 5: ...n using the 65 2434RK you must follow the instructions and warnings in this manual to assure proper and safe operation of the 65 2434RK and to minimize the risk of personal injury Be sure to maintain...

Страница 6: ...Filter The disc shaped charcoal filter is secured to the sensing face of the sensor by a rubber boot The filter prevents interference gases hydrogen sulfide H2S and certain hydrocarbons from producing...

Страница 7: ...null potentiometer It is factory set Do not adjust it Test Points The test points labeled CAL and CAL are on the right of the interconnect terminal strip The test points produce a 100 500 mV output th...

Страница 8: ...maintenance and calibration procedures Select a site where the target gas is likely to be found first NOTE If your application does not require a specific mounting site mount the transmitter at appro...

Страница 9: ...terminal labeled FB CAUTION Leave the shield drain wire insulated and disconnected at the transmitter You will connect the opposite end of the cable s drain wire at the controller 6 Secure the cover...

Страница 10: ...necessary to apply zero air when verifying or setting the fresh air reading The procedure below describes applying zero emission air usually called zero air using a calibration kit that includes a ca...

Страница 11: ...dure describes a test to verify that the CO transmitter responds properly to carbon monoxide NOTE Performing a response test on the CO transmitter may cause alarms Be sure to put the controller into i...

Страница 12: ...ion kit as described in the Calibration section of this insert 7 When the display reading falls below the alarm setpoints return the controller to normal operation Quarterly Calibrate the CO transmitt...

Страница 13: ...er instruction Replacing Components of the CO Transmitter This section includes procedures to replace the CO sensor amplifier PCB and charcoal filter Replacing the Sensor 1 Turn off the controller 2 T...

Страница 14: ...y plug the CO sensor with rubber boot and charcoal filter into the socket pattern on the amplifier PCB NOTE Match the sensor s male pins with the four female sockets as you plug the sensor into the so...

Страница 15: ...present and calibration adjustments are minimal at calibration then a calibration frequency of every 6 months is adequate If an application is very demanding for example if CO is present often and in...

Страница 16: ...ator s on off knob clockwise to close it 9 Remove the calibration cup from the sensor face 10 Unscrew the regulator from the zero air calibration cylinder 11 Leave the sample tubing connected to the r...

Страница 17: ...ts replacement parts and accessories for the CO channel Table 3 Parts List Part Number Description 06 1248RK Sample tubing order by the foot 07 0203RK Retaining boot for filter 33 7101RK Filter charco...

Отзывы: