65-2405RK-04 Combustible Gas Transmitter • 5
Figure 2: Mounting the Combustible Gas Transmitter
2.
At the monitoring site you select, hang or mount the detector junction box with the
detector facing down (see Figure 2).
3.
Install 3/16 inch I.D. flexible polyurethane tubing to the fitting at the bottom of the
calibration cup/splash guard and route it to an accessible area that is close to the
amplifier junction box. See the “Parts List” on page 19 for available tubing.
Rubber
Spacer,
3X
2.70
.38
J-Box
Combustible
Gas Detector
3.65
3/4 NPT
Female
Calibration
Cup/Splash
Guard
10.30 MAX
5.25
1.13
www.
.com
1.800.561.8187