20 • 65-2322RK Oxygen Transmitter
Returning to Normal Operation
1. Remove the voltmeter leads from the amplifier test points.
2. Secure the junction box cover to the junction box.
3. When the controller display reading rises above the decreasing alarm setpoint, return the
controller to normal operation.
NOTE:
If you do not allow the gas reading to increase above the decreasing alarm point, then
unwanted alarms may occur.
4. Verify that the controller display reading increases and stabilizes at 20.9% volume.
5. Store the components of the calibration kit in a safe and convenient place.