Overview

This manual describes the 65-2301RKSS toxic detector. This manual also describes how to install, start 
up, maintain, and calibrate the toxic detector when used with a gas monitoring controller. A parts list 
at the end of this manual lists replacement parts and accessories for the toxic detector.

The 65-2301RKSS detector can be used for various target gases which are listed in Table 1 below. For a 
particular target gas, the part number of the toxic detector will be of the format 65-2301RKSS-XXX. For 
example, 65-2301RKSS-NH3 is for ammonia (NH

3

) detection.

Specifications

WARNING: Do not use this product in a manner not specified in this instruction manual.

Table 1 lists specifications for the various 65-2301RKSS toxic detectors.

WARNING: When using the 65-2301RKSS, you must follow the instructions and warnings in this 

manual to assure proper and safe operation of the 65-2301RKSS and to minimize the 
risk of personal injury. Be sure to maintain and periodically calibrate the 65-2301RKSS 
as described in this manual.

Table 1: Specifications

Target Gas

65-2301RKSS-ASH3: Arsine (AsH

3

)

65-2301RKSS-CL2: Chlorine (Cl

2

)

65-2301RKSSCL-10: Chlorine (Cl

2

)

65-2301RKSS-HCN: Hydrogen Cyanide (HCN)
65-2301RKSS-NH3: Ammonia (NH

3

)

65-2301RKSS-SO2: Sulphur Dioxide (SO

2

)

65-2301RKSS-PH3: Phosphine (PH

3

)

Sampling Method

Diffusion

Detection Range

AsH

3

: 0 - 1.50 ppm

Cl

2

: 0 - 3.00 ppm

Cl

2

: 0 - 10.0 ppm

HCN: 0 - 15.0 ppm
NH

3

: 0 - 75.0 ppm

SO

2

: 0 - 6.00 ppm

PH

3

: 0 - 1.00 ppm

Accuracy

± 10% of reading or ± 5% of full scale (whichever is greater)

Response Time

• Cl

2

 and NH

3

: T90 in 90 seconds

• All others: T90 in 45 seconds

Operating Temperature & 

Humidity

• 14°F to 104°F (-10°C to 40°C)
• 20% to 90% Relative Humidity

Содержание 65-2301RKSS Series

Страница 1: ...www rkiinstruments com 65 2301RKSS Toxic Detector Operator s Manual Part Number 71 0273RK Revision A Released 7 17 19...

Страница 2: ...alibration and maintenance of the detector is essential for proper operation and correct readings Please calibrate and maintain this detector regularly Frequency of calibration depends upon the type o...

Страница 3: ...t costs local repair costs transportation costs or contingent expenses incurred without our prior approval THIS WARRANTY IS EXPRESSLY IN LIEU OF ANY AND ALL OTHER WARRANTIES AND REPRESENTATIONS EXPRES...

Страница 4: ...the Toxic Detector to a Controller 8 Startup 10 Introducing Incoming Power 10 Setting the Zero Fresh Air Signal 10 Maintenance 10 Preventive Maintenance 10 Troubleshooting 11 Replacing Components of...

Страница 5: ...2301RKSS toxic detectors WARNING When using the 65 2301RKSS you must follow the instructions and warnings in this manual to assure proper and safe operation of the 65 2301RKSS and to minimize the ris...

Страница 6: ...65 2300RK Toxic Detector The 65 2300RK toxic detector consists of the detector housing body detector housing cap splashguard cap gasket and the plug in sensor Figure 2 65 2300RK Toxic Detector Compone...

Страница 7: ...as Through a series of chemical and electrical reactions the sensor produces an electrical output that corresponds to the detection range of the sensor Junction Box The stainless steel corrosion resis...

Страница 8: ...g the Toxic Detector to a Controller WARNING Always verify that power to the controller is OFF before you make wiring connections 1 Turn off the controller 2 Turn off or unplug power to the controller...

Страница 9: ...rough the same conduit hub The power cable may disrupt the transmission of the detector signal to the controller 8 Connect the wires to the applicable controller terminal strip See the controller oper...

Страница 10: ...ccurate zero setting See Calibration on page 14 for instructions to introduce zero air to the detector 1 Verify that the detector is in a fresh air environment environment known to be free of toxic an...

Страница 11: ...r in this section If the fail condition continues contact RKI for further instruction Slow or No Response Difficult or Unable to Calibrate Symptoms Unable to accurately set the zero or response readin...

Страница 12: ...move the toxic sensor 5 Remove the black plug from the top of the replacement sensor 6 Carefully plug the replacement sensor into the connector that is located in the detector housing body WARNING You...

Страница 13: ...llow the replacement detector to warm up for 5 minutes before you continue with the next step 11 Calibrate the replacement detector as described in Calibration on page 14 Calibration Frequency Althoug...

Страница 14: ...ke sure to connect the sample tubing to the calibration cup port that results in the specified flow direction for the 81 1138RK calibration cup Preparing for Calibration 1 Unscrew the splashguard from...

Страница 15: ...nder and toxic gas cylinder If necessary to fit the calibration toxic gas cylinder change the regulator Setting the Response Reading Span 1 Screw the regulator into the calibration gas cylinder 2 Foll...

Страница 16: ...300RK SO2 Replacement detector assembly SO2 includes plug in sensor 71 0273RK 65 2301RKSS Operator s Manual this document 81 0076RK Zero air calibration cylinder 17 liter 81 0076RK 01 Zero air calibra...

Страница 17: ...on cup and a 34 liter 10 ppm NH3 in nitrogen aluminum calibration cylinder 81 F603RK Calibration kit includes regulator calibration cup and 58 liter 2 ppm CL2 in nitrogen aluminum calibration cylinder...

Страница 18: ...18 65 2301RKSS Toxic Detector ESM K01D CL2 10 ESM 01 plug in sensor 0 10 0 ppm chlorine diffusion type only Table 3 Parts List Part Number Description...

Отзывы: