35-3010A-11 Sample-Draw Detector • 5
Overview
This manual describes the 35-3010A-11 sample draw detector. This manual also describes
how to install, start up, maintain, and calibrate the sample draw detector when it is used
with a gas monitoring controller. A parts list at the end of this manual lists replacement
parts and accessories for the detector.
Specifications
Table 1 lists specifications for the 35-3010A-11.
WARNING: When using the 35-3010A-11, you must follow the instructions and
warnings in this manual to assure proper and safe operation of the
35-3010A-11 and to minimize the risk of personal injury. Be sure to
maintain and periodically calibrate the 35-3010A-11 as described in this
manual.
Table 1: Specifications
Target Gases & Detection Ranges
Combustible Gas: 0 - 100% LEL
Hydrogen Sulfide (H
2
S): 0 - 100 ppm
Carbon Dioxide (CO
2
): 0 - 5% volume
Input Power
24 VDC
Current Draw
200 mA
Construction (housing)
Fiberglass/polyester (NEMA 4X)
Dimensions
15.44 in. H x 12.55 in. W x 8.31 in. D
Weight
14 lbs.
Sampling Method
Sample draw
Sample Flow
1.2 SCFH (nominal)
Response Time
90% in 30 seconds
Accuracy
Combustible Gas:
± 5% of reading or ± 2% LEL (whichever is greater)
Hydrogen Sulfide:
± 5% of reading or ± 2 ppm H
2
S (whichever is greater)
IR Carbon Dioxide (CO
2
):
± 5% of reading or ± 2% of full scale (whichever is greater)