22 • 35-3001-06H-DIL Hydrogen/Oxygen Sample-Draw Detector
7. Connect the sample tubing from the regulator to the inlet line.
8. Allow the sample-draw detector to draw the calibrating sample for 1 minute.
9. Follow the directions in the controller’s operator’s manual for setting the zero (oxygen free)
reading.
10. Disconnect the gas bag tubing from the sample-draw detector’s inlet fitting.
11. Empty the gas bag.
12. Reconnect the incoming sample line to the inlet line.
13. Unscrew the fixed flow regulator from the calibration cylinder.
Returning to Normal Operation
1. Wait 1 to 2 minutes to allow the oxygen reading at the controller to return to normal, then
return the controller to normal operation.
NOTE:
If you do not allow the oxygen reading to return to normal, then unwanted alarms may
occur.
2. Verify that the controller display reading stabilizes at 20.9%.
3. Store the components of the calibration kit in a safe and convenient place.