background image

 

6 • 35-3001-06H-DIL Hydrogen/Oxygen Sample-Draw Detector

half of what the flowmeter indicates. The flowmeter measures the flow in the range 0.2 to 2.0 
SCFH (Standard Cubic Feet per Hour). The optimum flow rate is 0.8 SCFH.

Sensor Flow Control Valve

The sensor flow control valve is mounted to the flowmeter circuit board above the flowmeter. The 
sensor flow control valve adjusts the flow rate to the detectors. Turn the valve’s knob clockwise to 
increase the flow and counterclockwise to decrease the flow.

Status LEDs

Two status LEDs are above the flowmeter. They are also visible through the window in the housing 
door. The green Pilot LED is on when the sample-draw detector is receiving power from the 
controller. The red Fail LED is on when the sample flow rate is below the low flow level.

Pressure Switch

The pressure switch is mounted to the back of the flowmeter circuit board. The pressure switch 
monitors the flow rate of the incoming gas sample.

If the flow rate falls below the preset low flow level, the pressure switch causes the Fail LED to 
turn on and interrupts the signal from the detector. The interrupted detector signal causes a fail 
condition at the controller. The low flow level is factory-set at 1.0 SCFH (±0.1 SCFH). See 
“Adjusting the Low Flow Setting” on page 17 to adjust this setting.  

Pump Reset Switch

The pump reset switch is located to the left of the status LEDs. When a low flow condition occurs, 
the pump will be shut off. To reset the low flow condition and start the pump again, press and hold 
the pump reset switch for about 2 seconds, then release. 

Hydrophobic Filter

The hydrophobic filter is located toward the bottom left of the main circuit board. The filter 
prevents particulates and water in the incoming gas sample from damaging the flow and detection 
systems. Replace the filter when it appears dirty, discolored, or clogged.

Oxygen Sensor

The oxygen sensor is installed in a flow block and the flow block is mounted to the middle of the 
main circuit board. It is the upper of the two flow blocks mounted to the main circuit board. The 
oxygen sensor is retained in the flow block by a bracket with two screws.

The oxygen cell is protected within the sensor assembly. Through a series of chemical and 
electronic reactions, the oxygen cell produces a millivolt output that is proportional to the detection 
range of the sample-draw detector. The leads extending from the sensor terminate in lugs that 
connect to the oxygen terminal strip.

Dilution Fitting

The catalytic hydrogen sensor requires oxygen to operate. In environments where there is not 
enough oxygen to operate the hydrogen sensor, the dilution fitting adds sufficient oxygen by 
blending ambient air with the incoming sample. The standard dilution fitting dilutes at a ratio of 
approximately 1:1 (one part air to one part sample).

Hydrogen Sensor

The catalytic hydrogen sensor detects hydrogen in the %LEL range. It uses a catalytic element for 
detection. The reaction of gas with oxygen on the catalyst causes a change in the resistance of the 
element which changes the current flowing through it. The current is proportional to the detection 
range of the sample-draw detector.

The hydrogen sensor is installed in the flow block and the flow block is mounted to the middle of 
the main circuit board. It is the lower of the two flow blocks mounted to the main circuit board. An 

Содержание 35-3001-06H-DIL

Страница 1: ...www rkiinstruments com 35 3001 06H DIL Hydrogen Oxygen Sample Draw Detector Operator s Manual Part Number 71 0346 Revision A Released 8 7 17...

Страница 2: ...th Periodic calibration and maintenance of the detector is essential for proper operation and correct readings Please calibrate and maintain this detector regularly Frequency of calibration depends up...

Страница 3: ...or replacement costs local repair costs transportation costs or contingent expenses incurred without our prior approval THIS WARRANTY IS EXPRESSLY IN LIEU OF ANY AND ALL OTHER WARRANTIES AND REPRESENT...

Страница 4: ...g the Zero Reading 12 Maintenance 13 Preventive Maintenance 13 Troubleshooting 14 Replacing Components of the Sample Draw Detector 16 Adjusting the Low Flow Setting 17 Removing the Particle Filter s T...

Страница 5: ...sample through the dilution fitting Because of this it is important when calibrating the detector head to apply the calibration gas to the sample line inlet if a sample line is installed In addition b...

Страница 6: ...001 06H DIL and to minimize the risk of personal injury Be sure to maintain and periodically calibrate the 35 3001 06H DIL as described in this manual Maximum Recommended Inlet Exhaust Line Length for...

Страница 7: ...ution Fitting LEL Flow Block LEL Sensor W AMP AMP 2 S Sensor Flow Control Valve Fail LED Pilot LED Pump Reset Switch Pressure Switch Adjustment Screw behind PCB Flowmeter W B G Interconnect Terminal S...

Страница 8: ...4 in rigid tubing See the Installation section on page 8 to connect tubing to the sample fittings Particle Filter A particle filter with a tubing stub on one end is shipped with the instrument but it...

Страница 9: ...t board See the Installation section on page 8 for all wiring procedures related to the sample draw detector Pump Connector The pump connector is the two point connector below the oxygen terminal stri...

Страница 10: ...hobic Filter The hydrophobic filter is located toward the bottom left of the main circuit board The filter prevents particulates and water in the incoming gas sample from damaging the flow and detecti...

Страница 11: ...to the detector On the other end five pins extend from the sensor and plug into the socket connector The connector allows you to replace the sensor without disconnecting the wiring Four leads extendin...

Страница 12: ...lect the mounting site Is there enough room to open the housing door and make wiring and tubing connections at the bottom of the housing Make sure there is sufficient room to perform start up maintena...

Страница 13: ...side of the particle filter and route it to the sampling area 1 4 O D rigid polypropylene Teflon or flexible polyurethane tubing may be used RKI Instruments Inc recommends using either 1 4 O D x 1 8 I...

Страница 14: ...ng and the inlet fitting 3 Hand tighten the nut on the exhaust fitting so the ferrules clamp on the tubing If the tube nut is removed see Figure 5 for the arrangement of the fitting components 4 Route...

Страница 15: ...lded cable is used connect the cable s drain wire to an available chassis earth ground at the controller RKI controllers typically have a ground stud that can be used to ground the cable s drain wire...

Страница 16: ...itch and plug the open end of the sample line with your thumb If the sensor flowmeter indication drops noticeably the instrument goes into a low flow alarm and the Fail LED turns on the incoming sampl...

Страница 17: ...hecks 1 Verify that the Pilot LED is on 2 Verify that the flowmeter indicates a flow rate of approximately 0 8 SCFH If necessary use the sensor flow control valve to adjust the flow rate to 0 8 SCFH T...

Страница 18: ...kit in a safe place 10 Exhale into the sample draw detector s inlet line 11 Stop exhaling into the sample line then verify that the oxygen reading at the controller decreased 12 If the reading does no...

Страница 19: ...inues contact RKI Instruments Inc for further instruction Slow or No Response Difficult or Unable to Calibrate Symptoms The sensor responds slowly or does not respond during the monthly response test...

Страница 20: ...or to warm up for 5 minutes before you continue 12 Calibrate the replacement sensor as described in Calibration LEL Sensor on page 19 Replacing the Oxygen Sensor 1 Turn off the controller 2 Turn off p...

Страница 21: ...tub 3 Install the new particle filter onto the tubing stub Be sure the arrow on the particle filter is pointing toward the inlet fitting 4 Reinstall the tubing routed to the sampling area 4 If the par...

Страница 22: ...n the reset switch is pressed and released 5 Decrease the flow very slowly using the inlet line flowmeter s adjustment valve and verify that the low flow alarm is 1 0 SCFH 0 1 at the inlet line flowme...

Страница 23: ...te for most sample draw detector applications Unless experience in a particular application dictates otherwise RKI Instruments Inc recommends a calibration frequency of every 3 months If an applicatio...

Страница 24: ...egulator from the zero air calibration cylinder Setting the Response Reading 1 Screw the fixed flow regulator into the hydrogen calibration cylinder 2 Attach the gas bag tubing to the regulator on the...

Страница 25: ...on the zero air calibration cylinder NOTE A demand flow regulator cannot be used for calibration 2 Turn the regulator s on off knob clockwise to open it Fill the gas bag 3 Turn the regulator s on off...

Страница 26: ...r s inlet fitting 11 Empty the gas bag 12 Reconnect the incoming sample line to the inlet line 13 Unscrew the fixed flow regulator from the calibration cylinder Returning to Normal Operation 1 Wait 1...

Страница 27: ...rticle filter 61 0155RK Hydrogen sensor plug in type LEL range 71 0346 Operator s Manual 35 3001 06H DIL Sample Draw Detector 81 0002RK 01 Calibration cylinder 50 LEL hydrogen in air 34 liter steel 81...

Отзывы: