(4)  This instrument has no power supply switch nor fuses. Therefore,

install the fuse close to the instrument and the switch, if required.
Fuse type :

Time-lag fuse

Recommended fuse rating : Rated voltage : 250 V

Rated current : 1 A

(5) Conduct instrument power wiring so as not to be influenced by

noise from the electric equipment power. If the instrument may be
affected by external noise, a noise filter should be used.

y

 Shorten the distance between twisted power supply wire pitches.

The shorter the distance between the pitches, the more effective for
noise reduction.

y

Install the noise filter on the panel which is always grounded and
minimize the wiring distance between the noise filter output side
and the instrument power terminals.

     

OUT

Minimize
distance

power

terminals

IN

Twist these leadwires

Shorten distance between

pitches

Instrument power

Noise filter

y

Do not install fuses and/or switches on the filter output signal since
this may lessen filter effect.

3.2 Solderless terminal

Space of fitting a solderless terminal is 7.4 mm

×

11.3 mm. Therefore,

use the solderless terminal suitable for a screw of M3. A free terminal

screw of M3

×

6 is used.

When the instrument is installed in a location where much
vibration and shock exist, use ring type compression lugs,
since spade type compression lugs may be desconnected from
the terminals.

4. CONNECTION

4.1 Connection method

This device supports communications between the RS-232C and RS-

422A or RS-485 interfaces.

 

Connection procedure of host computer

Host computer and COM-103C are connected by a cable for RS-

232C (Using the connector).

c

 Loosen the plug case screws, connect the cable core wires to the

relevant connector pins, and mate the plug case with the wired

connector before tightening the screws with a screwdriver.

d

 

Plug the communication connector connected with the cable 

c

 into

the instrument connector (See the terminal plate diagram.)

e

 Then, using a screwdriver, secure the connector with 2 upper and

lower screws.

Communication cable for RS-232C

Host Computer

COM-103C

SD

RD

RS

CS

SG

SG

CS

RS

RD

SD

2

3

4

5

7

2

3

4

5

7

Accessories

Connector 

HDBB-25P

 (Hirose Electric Co., Ltd. or equivalent)

Plug case 

HDB-CTF (

Hirose Electric Co., Ltd. or equivalent

)

 

Connection procedure of controller

COM-103C and controller are connected by a twisted pair wire
(shielded). (Using the terminals #1, #2, #3 of COM-103C.)
The twisted pair wire should be prepared by the user.
For details of connection procedure, see the communications
instruction manual for using RKC

s controller.

R K C

s

c o ntro lle r

R S -42 2 A   or  R S -4 85

S G

T/R (A )

T/R (B )

Tw is ted   p air  w ire
(w ith  s hie ld)

C O M -1 03 C

R S -23 2 C

Tw is ted   p air  w ire
(w ith  s hie ld)

H o s t  C o m pu te r

1

2

3

9

S G

F G

T/R (A )

T/R (B )

*R

*R   :  U s e  a   term ina l  res is to r  w ith   a   c o m b ine d  re s is tan c e   o f  m o re   th an   1 00  

.

P a ire d  w ire

4.2   Multidrop connection example

If branched with our instrument BRA-100B, the transmission distance

between COM-103C and BRA-100B may be extended to some degree

and multidrop connection becomes possible. The number of controllers

that can be connected depends on the controller being used.

For details of RKC

s BRA-100B, see the instruction manual.

Host Computer

RS-232C

COM-103C

BRA-100B

BRA-100B

BRA-100B

RS-422A or RS-485

RKC

s

controller

RKC

s

controller

RKC

s

controller

RKC

s

controller

1

4

2

3

1

4

2

3

3

2

1

4

  3

 

1

 

2

 

4

29

30

31

   

4.3   Internal Connection

R S -23 2C

H os t  C o m p uter

C O M -1 03 C

S D

R D

R S

C S

S G

 

3

 

2

 

5

 

7

R D

S D

C S

S G

T /R (A )

(te rm ina l N o.2)

T /R (B )

(te rm ina l N o.3)

S G

(te rm ina l N o.1)

R S -42 2A

or

R S -48 5

*

*ON : Output the data of RS-232C side to RS-422A (RS-485) side.

  OFF : Output the date of

 

RS-422A (RS-485) side to RS-232C side.

CAUTION

HEADQUARTERS: 16-6, KUGAHARA 5-CHOME, OHTA-KU TOKYO 146-8515 JAPAN

PHONE

IMCOM04-E3

E-mail

JUN.1999

FAX

: 03-3751-9799 (+81 3 3751 9799)

[email protected]

: 03-3751-8585 (+81 3 3751 8585)

RKC INSTRUMENT INC.

Отзывы: