11.Detection principle
11-1
.
NDIR method (Non-Dispersive Infrared)
The model is based on NDIR method (Non-Dispersive Infrared) and this structure is shown below.
The infrared beam emitted from the light source passes through the measuring cell, and optical
band pass filter which can pass the absorption wave of measuring gas and attains to the infrared
sensor. The amount of infrared attaining to the infrared sensor through the measuring cell and
will decrease according to its density.
The variable amount of infrared is measured by the infrared sensor and it is displayed as gas
concentration.
Then, there is no sensitivity against CO etc which have the different absorption wave from the
measuring gas. Then there is no sensitivity against N2 and H2 etc which cannot absorb infrared.
As compared with the conventional catalytic combustion method, there is no poisoning material
to be absorbed and almost no sensitivity drop on this detection principle.
チョ ッパ
セル
光源
赤外 セン サ
光学 フィ ルター
構 造 図
11-2.
GALVANIC CELL method
Galvanic cell is consisting of a lead anode and a gold cathode in electrolyte covered by a
membrane. When oxygen enters into the gold cathode, a current which is directly proportional
to the oxygen concentration will be produced, and amplified current will produce a reading on
a meter in percent oxygen.
隔 膜
陰 極
電 解 液
サ ー ミス タ
陽 極
出 力 端子
構 造 図
Light source
Measuring cell
Photo filter
Figure
Membrane
Lead electrode
(Anode)
Gold electrode
(Cathode)
Solution
Output terminal
Thermistor
Figure