<FLOW CHART OF ZERO POINT ADJ>
Display
HOLD KEY
・Press ZERO key for 3 sec.
ZERO key
Mount CO2 scrubber
FRESH AIR?
ENTER key
ESC key
Move to zero adj
for CO2 adj ESC
key
ADJ ZERO
【Zero set OK】
ZERO SET OK
FAIL ZERO
【ZERO adj fails】
S C R U B B E R?
ENTER key
ADJ ZERO
Go back to the operation
display by ESC key.
ESC key
《ZERO sets OK》
ZERO SET OK
Press for 5 sec and move to
zero adj of CO2
・When press ENTER key, CH4, O2, CO zero cal
starts
・Calibration
Gas display
All are OK
CH
4
fails, CO
2
fails
Shows 5 sec and return back to gas
density screen
*Note
To cancel previous operation, press ESC key.
But can not cancel after press ENTER key.
《ZERO fails》
・When press ENTER key, zero adj of
CO2 starts
FAIL ZERO