3 How to Connect Wire
3-1. Description of terminal plate
- 19 -
M4 is used as the terminal screws for the terminal plate. Attach an insulated ring terminal with insulating
coating for M4 to the tip of a cable for wiring.
NOTE
If you are considering the use of the RS-485 communication function, contact RIKEN KEIKI.
NOTE
Contact output 3 (SSR)
•
Contact output 3 of the calorimeter makes SSR output.
•
AC power supply is required to drive the contact output.
Converting SSR contact into dry contact
Add an external relay.
Contact output 3 of the calorimeter makes SSR output. (Load specification: 75 - 264 VAC, 20 mA - 1 A)
It is not possible for a DC input unit to obtain SSR output.
In this case, add an external AC relay to convert SSR contact output into dry contact output.
<Wiring Diagram>
•
Install an AC relay in the non-hazardous area.
•
Use a coil with specified voltage of 75 - 264 VAC and current of 20 mA - 1 A for the relay.
E.g.) MK3ZP (OMRON): Rated current 23.1 mA (100 VAC, 60 Hz)
OHC-800
(5) (6)
Relay
~
~
100 VAC
Input unit
Hazardous location
Non-hazardous location
Содержание RS-400 Series
Страница 1: ...PT3E 0512 Explosion Proof Calorimeter OHC 800 Sampling Device RS 400 Series Operating Manual...
Страница 6: ......
Страница 7: ......
Страница 8: ......
Страница 10: ......
Страница 11: ......
Страница 12: ......
Страница 14: ......
Страница 15: ......
Страница 16: ......
Страница 19: ......
Страница 20: ......
Страница 23: ......
Страница 24: ......
Страница 25: ......
Страница 28: ......
Страница 30: ......
Страница 33: ......
Страница 34: ......
Страница 35: ......
Страница 36: ......
Страница 37: ......
Страница 38: ......
Страница 40: ......
Страница 41: ......
Страница 42: ......