5 How to Operate
5-3. How to start the pump unit
- 31 -
RP-D58 Series
5-3. How to start the pump unit
Before supplying power to the pump unit, check that the preparation for start-up is completed.
Open the electric circuit lid.*
Turn ON the power switch.*
Close the electric circuit lid.*
Supply power to the pump unit.
After the pump unit completes the start-up, it enters the detection mode swiftly.
* Only RP-D58·AC has a power switch. Since RP-D58·DC does not have any power switch, there is no
need to open or close the electric circuit lid.
NOTE
<<Start-up Operation Procedures>> (approximately 25 seconds for system check of the pump unit and
alarm deactivation)
Power on -> Initial clear (approximately 25 seconds) -> Detection mode
The gas detection is not yet started just after the pump unit is started up. Refer to see the operating
manual of the indicator/alarm unit and gas detector head.
DANGER
Do not turn off the pump unit during the initial clear. The pump unit is reading the internal memory
during the initial clear.
Power-on
Initial clear
Detection mode