43
11. Specifications
11-1 Standard specification
Model
:
RI-2000W
Detecting principal
:
Non-dispersive infrared absorption (NDIR)
Target gas
:
Nitrous Oxide(N
2
0)
Detection range
:
0 to 200 ppm
Alarm accuracy
:
Within 30% of the alarm set point (on a like-for-like basis)
Alarm delay time
:
Within 30 seconds
Sampling method
:
Automatic aspiration by a pump
Aspiration rate
:
approximately 1.0 litre/ min
Gas inlet connector
:
φ
6(0D)-
φ
4(1D) size Teflon tube with a propylene male
connector
Indications
:
16 digits (numbers and alphabets)
Concentration
:
LCD 16 digits
Status
:
LCD 16 digits and LED lamps
1.
POWER
lamp
(Green)
・
Normal condition
:
Keeps on
・
Trouble condition
:
Flashes
2.WARNING
lamp
(Yellow)
・
Normal condition
:
Off
・
1st gas alarm
:
Flashes
(
Keeps on if BZ.STOP is pressed
)
3.ALARM lamp
(Red)
・
Normal condition
:
Off
・
2nd gas alarm
:
Flashes
(
Keeps on if BZ.STOP is pressed
)
4.MAINTENANCE lamp
(Green)
・
Detecting mode
:
Off
・
Maintenance mode
:
Flashes
transmission scheme
:
4-20mA (non-insulated, discharge energy)
load resistance
:
below 300
Ω
Output signal
:
4-20 mA
Warm up
:
2.5 mA
Maintenance
:
2.5 mA
Trouble (fault)
:
0.5 mA
Recommended cables
:
CVVS
relevant cables 1.25 to 2.0 sq