5 Operation Procedures
5-1 Preparation
Please confirm these things written below before starting the instrument. It will cause
damage in the instrument or cause electrical shocks if there is a lack of preparation.
1.
Is it earthed properly?
2.
Is it the wiring done properly?
3.
Is the power source within the rating?
4.
Is the output contact not able to affect other instruments? The output contact may
activate during the adjustment. Please make sure the output contact cannot effect
on others.
5.
Is the dust filter mounted?
6.
Is the fuse within the rating? If not, it may cause fire.
5-2 Basic motion flow chart
Usually, please operate this instrument in “Detection mode”.
*
NOTE
Please let RI-2000W to operate at least 4 hours warm up after it hasn’t been used for
a long time.
Start
Preparation
Warm up (60minutes)
Detection
mode
Zero calibration mode
press and hold down ZERO/
▲
for 3 sec.
Alarm test mode
press and hold down TEST/
▼
for 3 sec.
Check alarm point
press and hold down ALARM/SET for 3 sec.