5.4 Powering off the GX-3R / GX-3R Pro
【
In the case of SM
】
1. Hold down the POWER + EDIT / ENTER switch for 3 seconds to turn off the GX-3R / GX-3R Pro.
* This operation can be performed only on the basic screen and result screen of 4.3.
2. If there is no switch operation for 10 minutes or more on the basic screen or result screen, the power of
the GX-3R / GX-3R Pro is turned off.
3. Even if the power of the GX-3R / GX-3R Pro is turned off, the bump test / gas calibration / alarm check
result (LED) state is maintained.
4. After that, turn on the power of GX-3R / GX-3R Pro and display the result of the previous bump test /
gas calibration / alarm check.
※
The previous result display is only when the GX-3R / GX-3R Pro of the same serial number is
connected.
* The result is cleared when the power of SDM-3R is turned off, so the previous result is not displayed
even if the
power of GX-3R / GX-3R Pro is turned on.
※
If the GX-3R / GX-3R Pro with different serial number is connected, the result of bump test /
gas calibration / alarm check will be cleared and the display will be the basic screen (see 4.3).
【
In the case of DM
】
1. In the PC software, select "Power off" from the right-click menu of the detector icon to turn off the
GX-3R / GX-3R Pro.
2. If there is no operation for 1 hour on the main screen, the power of the GX-3R / GX-3R Pro will be
turned off. (Time can be changed)
CH4 %LEL O2
%
P P
P P
CO
ppm
H2S
ppm
P P
F P
BUMP/CAL
CH4 %LEL CO ppm H2S ppm
P P
P P
F P
O2
vol%
CO2
ppm
P P
P
BUMP / CAL