6 Alarm Activation and Functions
6-3. External output operation
- 38 -
6-3. External output operation
<Range:0-25vol% L-LL, L-H 4 - 20 mA Output Table>
Status
External output
(4 - 20 mA)
External output
(0 - 1 V)
Remarks
Initial
17.4mA
0.84V
Fixed values
Normal
4.0 - 20.0mA
0 - 1V
Depending on gas
concentration
Scale over
20.1mA -
1V
Depending on gas
concentration
Trouble
21.5mA
1V
Fixed values
Maintenance mode
menu
17.4mA
0.84V
Fixed values
Air calibration
17.4mA
0.84V
Fixed values
Alarm setpoint check
17.4mA
0.84V
Fixed values
Alarm summary
check
17.4mA
0.84V
Fixed values
Date/time check and
setting
17.4mA
0.84V
Fixed values
Alarm test
4.0 - 20.0 mA
0 - 1V
Depending on test
value
<Range:0-50vol% L-LL, L-H 4 - 20 mA Output Table>
Status
External output
(4 - 20 mA)
External output
(0 - 1 V)
Remarks
Initial
10.7mA
0.42V
Fixed values
Normal
4.0 - 20.0mA
0 - 1V
Depending on gas
concentration
Scale over
20.1mA -
1V
Depending on gas
concentration
Trouble
21.5mA
1V
Fixed values
Maintenance mode
menu
10.7mA
0.42V
Fixed values
Air calibration
10.7mA
0.42V
Fixed values
Alarm setpoint check
10.7mA
0.42V
Fixed values
Alarm summary
check
10.7mA
0.42V
Fixed values
Date/time check and
setting
10.7mA
0.42V
Fixed values
Alarm test
4.0 - 20.0 mA
0 - 1V
Depending on test
value