5 Operations and Functions
5-2. Fault alarm activation
GP-1000
- 48 -
5-2. Fault alarm activation
Fault alarm:
Triggered when an abnormality is detected in the gas detector.
Fault alarm activation:
Self-latching
Alarm display:
Notifies by lighting of the alarm lamp, sounding of the buzzer and displaying the
detail.
Alarm types:
System abnormalities, sensor abnormalities, calibration abnormalities, low
battery voltage, low flow rate and clock abnormalities
<Display Operation>
Alarm type
System abnormalities, sensor abnormalities, calibration abnormalities, low
battery voltage, low flow rate and clock abnormalities
Alarm lamp
Blinks at about 1 second intervals.
Buzzer
Sounds intermittently at about 1 second intervals.
LCD display
Displays an error message
NOTE
Of the fault alarms, only the low flow rate alarm "FAIL - LOW FLOW" and "FAIL - CLOCK" can be
reset by pressing the RESET button. For other fault alarms, turn off the power and then promptly
contact RIKEN KEIKI.
For information on malfunctions (error messages), see "8. Troubleshooting".
Display example (low flow rate)