4 How to Use
4-8. Calibration mode
GP-1000
- 38 -
4-8-2. Bump test
The function is tested using a test gas. The result will be displayed as "P" (PASS) or "F" (Failure). If the
function is diagnosed as "F" (Failure), take a measure such as performing span calibration.
LCD
Display
Bump test
(1) Press the ENTER button to
enter the bump test.
[ENTER]
(2) Supply test gas and press the
ENTER button.
[ENTER]
* "BUMP" and "APPLY" are
displayed alternately and the
countdown is started. When the
count reaches zero, diagnosis is
performed.
→
←
↓
* When CAL is set to ON, "CAL"
and "APPLY" are displayed
alternately next and the
countdown is started. When the
count reaches zero, calibration is
performed.
→
←
↓
* The diagnosis result is displayed.
(3) Press the ENTER button to
return to the calibration mode
menu.
* P : PASS
F : Failure
[ENTER]