![Riken Keiki GD-F4A-A Скачать руководство пользователя страница 22](http://html1.mh-extra.com/html/riken-keiki/gd-f4a-a/gd-f4a-a_operating-manual_1461465022.webp)
20
9-4.Detection principle
[Detection principle]
A negative electrode of noble metal and a positive electrode of lead are placed in a resin container filled
with electrolyte. A part of the container is opened and covered with a barrier. The negative electrode is
installed in contact with the barrier. Lead wires are drawn from the positive and negative electrodes to
obtain outputs. A thermistor is connected between the lead wires to perform temperature compensation
for the sensor outputs.
Barrier
Electrolyte
Container
Lead wire
Thermistor
Output
Positive
electrode
Negative
electrode
Содержание GD-F4A-A
Страница 1: ...PT2E 2501 Oxygen Gas Detector Head GD F4A A Operating Manual PT2 151...
Страница 24: ...Manual Log Rev Amendment Issue data 0 First issue 2017 2 17 1 Addition Declaration of Conformity 2022 7 22...
Страница 25: ......
Страница 26: ...1...