Operating Precautions 

 
 
The detector detects oxygen in the atmosphere and prevents acid-deficient accidents due to acid-deficient 
air release or accidents due to excessive oxygen. 
Please fully understand the following points before using it, so that it can be used properly. 
 
1.  The detector may be interfered by gases and vapors other than oxygen in the atmosphere. 
 

In addition, it may be fluctuated by environmental (temperature, humidity etc.) changes in the 
installation site. 

 
2.  The detector is affected by pressure. If the detector is used at a pressure other than atmospheric 

pressure, a calibration is required under that pressure. The maximum allowable pressure is 110 KPa. 
The detector cannot be used under reduced pressure. 

 
3.  The gas sensor used by the detector contains an electrolyte. Electrolyte may cause severe skin burns if 

it contacts skin, while it may cause blindness if its contacts eyes. Do not touch the sensor directly in the 
event of a liquid leak. 

 
4.  For maintenance of the detector, it must go through a regular maintenance, including replacement and 

adjustment of the regular replacement parts as specified in the operating manual. In addition, because 
this is a safety unit, it is recommended that a regular maintenance and span adjustment are performed 
every six months. 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

Содержание GD-1DOXi

Страница 1: ...PT2E 2361 Fixed type Oxygen Detector Head GD 1OXi Series GD 1OXi GD 1DOXi Operating Manual PT2 236 PT2 237...

Страница 2: ...pressure other than atmospheric pressure a calibration is required under that pressure The maximum allowable pressure is 110 KPa The detector cannot be used under reduced pressure 3 The gas sensor use...

Страница 3: ...tem designing 11 4 4 How to install 12 4 5 Installation procedure 14 4 6 How to wire 16 4 7 Cable connection example 17 4 8 How to tube 19 5 How to Operate 20 5 1 Preparation for start up 20 5 2 Basic...

Страница 4: ...e oxygen sensor this compact diffusion type oxygen detector is used with an indicator alarm unit to measure a broad range of oxygen concentrations ranging from insufficient to excessive oxygen levels...

Страница 5: ......

Страница 6: ......

Страница 7: ......

Страница 8: ......

Страница 9: ......

Страница 10: ......

Страница 11: ......

Страница 12: ......

Страница 13: ......

Страница 14: ......

Страница 15: ...d by RIKEN KEIKI BARRIERTA JFE 552 manufactured by NOK KLUBER If you can not prepare the specified grease use one that meets the following requirements 1 Material does not harden due to deterioration...

Страница 16: ......

Страница 17: ......

Страница 18: ......

Страница 19: ......

Страница 20: ......

Страница 21: ......

Страница 22: ......

Страница 23: ......

Страница 24: ...tenance content Daily mainte nance Monthly mainte nance Regular mainte nance Power Supply Check Check that the power lamp lights up Status display Check Check that the fault lamp does not light up Con...

Страница 25: ...l on the oxygen deficiency meter by using the zero gas Performs the zero adjustment fresh air adjustment if the reading is incorrect Alarm Test Inspects the alarm circuit by using the alarm test funct...

Страница 26: ......

Страница 27: ......

Страница 28: ......

Страница 29: ......

Страница 30: ......

Страница 31: ......

Страница 32: ...s to indicator alarm unit as well See the operating manual of the indicator alarm unit for details Symptom Display FAULT Causes Actions Sensor error appears on the indicator side The wiring is not cor...

Страница 33: ...H 88 D mm projection portions excluded Weight Approx 2 5kg Color Munsell 7 5BG5 2 Recommended item Suction type Model GD 1DOXi Detection principle Galvanic cell method Detection gas O2 Detection metho...

Страница 34: ......

Страница 35: ...used to express the percentage of a specific substance or gas in a volume of solution Atmosphere An atmosphere with a temperature within the range of 10 to 40 C and a humidity of 95 RH or less at one...

Страница 36: ...Manual Log Rev Amendment Issue data 0 First issue 2018 12 19 1 P12 Corrected maintenance space 2020 6 29...

Отзывы: