2 Important Notices on Safety
2-3. Precautions
- 5 -
2-3. Precautions
CAUTION
Do not use a transceiver near the detector.
Radio wave from a transceiver or other radio wave transmitting device near the detector or its
cables may disturb readings. If a transceiver or other radio wave transmitting device is used, it
must be used in a place where it disturbs nothing.
Verify that the flow rate corresponds to the specified flow rate before using the detector.
If it does not correspond to the specified flow rate, gas detection cannot be performed properly.
Check whether the flow rate is unstable or lost.
Attach the dust filter before using the detector.
Before using the detector, attach the specified filter to prevent disturbances by possible gas
adsorption or air dust.
A dust filter to be used varies depending on the gas to be detected. For more information on dust
filters, please contact RIKEN KEIKI.
Observe the operating restrictions to prevent condensation inside the tube.
Condensation formed inside the tube causes clogging or gas adsorption, which may disturb
accurate gas detection. Thus, condensation must be avoided. In addition to the installation
environment, carefully monitor the temperature/humidity of the sampling point to prevent
condensation inside the tube. In particular, when detecting a gas which is dissolved into water and
corrodes contacted materials, such as a strong acid gas, the gas is undetectable and furthermore
may corrode internal parts. Please observe the operating restrictions.
Do not disassemble/modify the detector, or change the settings if not necessary.
Disassembling/modifying the detector will invalidate the warranty of the performance. Changing the
settings without understanding the specifications may cause alarm malfunctions. Please use the
detector properly in accordance with the operating manual.
Avoid applying organic solvents and others to the window plate for a long time.
The window plate material is a polycarbonate resin. When organic solvents (liquid or highly-
concentrated vapor) and others are applied to the plate for a long time, its color and shape may be
changed.
Never fail to perform a regular maintenance.
Since this is a safety unit, a regular maintenance must be performed to ensure safety. Continuing
to use the detector without performing a maintenance will deteriorate the sensitivity of the sensor,
thus resulting in inaccurate gas detection.
Содержание GD-1DOX-AS
Страница 1: ...PT2E 2880 Fixed type Oxygen Detector Head GD 1DOX AS Operating Manual PT2 183...
Страница 9: ......
Страница 10: ......
Страница 15: ......
Страница 16: ......
Страница 18: ......
Страница 22: ......
Страница 25: ......
Страница 32: ......
Страница 34: ...Manual Log Rev Amendment Issue data 0 First issue PT2 2880 1 31 2022...