6 Maintenance
6-1. Maintenance intervals and items
- 21 -
6
Maintenance
The detector is an important instrument for the purpose of safety.
To maintain the performance of the detector and improve the reliability of safety, perform a regular
maintenance.
Continuing to use the detector without performing maintenance will compromise the sensitivity of the gas
sensor, thus resulting in inaccurate detection.
6-1. Maintenance intervals and items
•
Daily maintenance: Perform maintenance before beginning to work.
•
Monthly maintenance: Perform maintenance on the alarm circuit (alarm test) once a month.
•
Regular maintenance: Perform maintenance once or more for every six months to maintain the
performance as a safety unit.
Maintenance
item
Maintenance content
Daily
mainte-
nance
Monthly
mainte-
nance
Regular
mainte-
nance
Power Supply
Check*
Check that the power lamp lights up.
○
○
○
Status display
Check*
Check that the fault lamp does not light up.
○
○
○
Flow rate check Check the flow rate to find abnormalities.
○
○
○
Concentration
Display Check*
Check that the concentration display value is zero (or
20.9 vol% on the oxygen meter). When the reading is
incorrect, perform the zero adjustment (or span
adjustment) after ensuring that no interference gases
exist around it.
○
○
○
Filter check
Check the dust filter for dust or clogging.
○
○
○
Alarm Test*
Inspect the alarm circuit by using the alarm test function.
-
○
○
Span
Adjustment*
Perform the span adjustment by using the calibration
gas.
-
-
○
Gas Alarm
Check*
Check the gas alarm by using the calibration gas.
-
-
○
* Check and adjustment are performed at the indicator/alarm unit side. See the operating manual of the
indicator/alarm unit for details.
NOTE
•
Before starting alarm test or span adjustment, be sure to provide a notification to the related
sections and block external signals.
Содержание GD-1DOX-AS
Страница 1: ...PT2E 2880 Fixed type Oxygen Detector Head GD 1DOX AS Operating Manual PT2 183...
Страница 9: ......
Страница 10: ......
Страница 15: ......
Страница 16: ......
Страница 18: ......
Страница 22: ......
Страница 25: ......
Страница 32: ......
Страница 34: ...Manual Log Rev Amendment Issue data 0 First issue PT2 2880 1 31 2022...