11
Compensation Adjustments
You are suggested to compensate the probe to match with the
oscilloscope. The probes can be adjusted for both low-frequency
compensation and high-frequency compensation. You should make
low-frequency compensation of the probe when it is connected to
the oscilloscope for the first time (or adjust it at any time), while
high-frequency compensation can be made regularly.
Low-frequency compensation
Follow the steps below to make low-frequency compensation:
1
.
Connect the probe to a channel BNC connector and the probe
compensation signal output terminal at the front-panel of the
oscilloscope. Connect the ground alligator lead of the probe to
the “Ground Terminal” under the probe compensation signal
output terminal.
2
.
Press Auto at the front panel of the oscilloscope.
3
.
Set the low-frequency compensation adjustment on the probe
using screwdriver provided via the low-frequency compensation
adjustment hole on the probe to get the flattest pulse possible
(as shown in figures on the next page).