CAUTION

1)

 

Please use function signal generator to output 20 Vp-p, 200 Hz square waveform. 

2)

 

Non-professionals do not adjust the Hole 2 and Hole 3. 

3)

 

This product is already adjusting Hole 2 and Hole 3. Please feel free to operation. 

4)

 

Use the adjust bar attached only. 

5)

 

When the measuring frequency was 40MHz up, you must replace the long earth 
lead with the Alligator Clip (BP-276N-D) to obtain a best earth and the best 
frequency response. 

 

 

5.

 

Warning

 

1)

 

Do not attempt to take the ground of test equipment away from the ground 
terminal. 
The ground connection is critical to the safety operation of the probe. Failure to 
make this connection may result in personal injury or damage to the probe or 
oscilloscope. You must be sure to connect the ground before the probe tip comes 
into contacting with the high voltage. And the ground connection must not be 
removed until the probe tip has been taken away from the high voltage source. 

2)

 

Do not connect the ground clip to the high voltage source or the probe tip to the 
ground for any reason. 

3)

 

Before turning the high voltage on, make sure that no part of your body is in 
contact with the device. 

4)

 

Remembering that the voltage being measured is 1000 times greater than the 

Adjust the frequency 
response of the probe to 
match your oscilloscope by 
the adjust bar attached only.   

For 40MHz up   
Measuring Frequency 

Hole 2.   

for 200kHz Square 

Wave Adjustment 

Hole 3.   
for Bandwidth 
Adjustment 

Hole 1. 
for 200Hz Square Wave Adjustment 

Содержание RP1018H

Страница 1: ...RIGOL User s Guide Publication Number UGE22X00 1110 RP1018H RP1018H High Voltage Probe Sept 2013 2013 RIGOL Technologies Inc All Rights Reserved ...

Страница 2: ...1 2 安全须知 1 3 操作说明 2 4 匹配电容调整 2 5 注意事项 3 6 清洁 4 7 电压测试临界曲线 4 8 附件 5 1 Specification 6 2 Safety Precautions 6 3 Operation 7 4 Compensation Adjustment 7 5 Warning 8 6 Cleaning 9 7 Voltage Derating Curve 9 8 Accessories 10 ...

Страница 3: ...RH or less at 35 Storage Temperature Range 20 70 Color Handel Body Black Yellow Weight Volume 460g 80 W x 80 H X 320 L mm 2 Safety Precautions Thank you for purchasing RP1018H high voltage probe High voltage probe can prevent high voltage testers from subjecting to unexpected shocks when doing high voltage measurement work Before using the high voltage probe the operator must read and fully unders...

Страница 4: ...e kept clean and free of any conductive contamination The probe is for indoor use only 3 Operation 1 Connect the divider probe common lead alligator clip to a good earth ground or reliable ground 2 Connect the BNC connector to the BNC input of your oscilloscope 3 Select the desired range of your oscilloscope NOTE Please turn the high voltage source off before making any connections 4 Compensation ...

Страница 5: ...ection may result in personal injury or damage to the probe or oscilloscope You must be sure to connect the ground before the probe tip comes into contacting with the high voltage And the ground connection must not be removed until the probe tip has been taken away from the high voltage source 2 Do not connect the ground clip to the high voltage source or the probe tip to the ground for any reason...

Страница 6: ...cotton cloth lightly moistened with alcohol and water Do not allow any portion of the probe to submerge in the liquid Dry the probe thoroughly before measuring voltage Do not place the probe into the corrosive solvents or fumes as these will cause deterioration to the probe and cables 7 Voltage Derating Curve RP1018H 18kV 3MHz 300V 150MHz 100V 1GHz Voltage dc ac peak Frequency ...

Страница 7: ...ccessories Test CRT for H V Adjust bar Carrying Box Contact Us If you have any problem or requirement when using our products or this manual please contact RIGOL E mail service rigol com Websites www rigol com ...

Отзывы: