RIGOL
2-2
User’s Guide for DSA1000 Series
2.1
Basic Settings
2.1.1
FREQ
Sets the frequency parameters of the analyzer. The analyzer sweeps within a specified
frequency range, and the sweep is restarted every time you change this parameter.
The frequency range of a channel can be expressed by either of two groups of
parameters: Start Frequency and Stop Frequency; or Center Frequency and Span. If
any such parameter is changed, the others would be adjusted automatically in order
to ensure the coupling relationship among them.
2
)
(
start
stop
center
f
f
f
+
=
(Equation 2-1)
start
stop
span
f
f
f
−
=
(Equation 2-2)
center
f
,
stop
f
,
start
f
and
span
f
denotes the center frequency, the stop frequency, the
start frequency and the span respectively.
2.1.1.1
Center Freq
Sets the center frequency of the sweep. When pressed, the frequency mode is
switched to Center Freq and Span in order to enter the desired parameter data. In this
mode, parameters you specified are always shown at the lower right and left side of
the display grid.
Key Points:
The start and stop frequencies vary with the center frequency when the span is
constant.
Changing the center frequency horizontally shifts the current sweep channel and
the adjustment is limited by the specified frequency range.
In Zero Span mode, the start frequency, stop frequency and center frequency are
always equal. If one is changed the others are updated to match.
You can modify this parameter using the numeric keys, knob, or direction keys.
Refer to “Parameter Input” for more details.
Содержание DSA1030
Страница 1: ...RIGOL User s Guide DSA1000 Series Spectrum Analyzer Sept 2014 RIGOL Technologies Inc...
Страница 2: ......
Страница 22: ......
Страница 179: ...Measurement Examples RIGOL User s Guide for DSA1000 Series 3 3 Figure 3 1 Measuring a Sine signal...
Страница 181: ...Measurement Examples RIGOL User s Guide for DSA1000 Series 3 5 Figure 3 2 Use the frequency counter to measure...
Страница 186: ...RIGOL Measurement Examples 3 10 User s Guide for DSA1000 Series Figure 3 4 Phase noise measurement...
Страница 188: ...RIGOL Measurement Examples 3 12 User s Guide for DSA1000 Series Figure 3 5 N dB bandwidth measurement result...
Страница 190: ...RIGOL Measurement Examples 3 14 User s Guide for DSA1000 Series Figure 3 6 Time domain power measurement...
Страница 192: ...RIGOL Measurement Examples 3 16 User s Guide for DSA1000 Series Figure 3 7 Adjacent channel power measurement result...
Страница 194: ...RIGOL Measurement Examples 3 18 User s Guide for DSA1000 Series Figure 3 8 Channel power measurement result...
Страница 196: ...RIGOL Measurement Examples 3 20 User s Guide for DSA1000 Series Figure 3 9 Occupied bandwidth measurement result...
Страница 198: ...RIGOL Measurement Examples 3 22 User s Guide for DSA1000 Series Figure 3 10 Emission bandwidth measurement result...
Страница 200: ...RIGOL Measurement Examples 3 24 User s Guide for DSA1000 Series Figure 3 11 C N Ratio measurement result...
Страница 204: ...RIGOL Measurement Examples 3 28 User s Guide for DSA1000 Series Figure 3 13 Harmonic Distortion measurement result...
Страница 211: ...Measurement Examples RIGOL User s Guide for DSA1000 Series 3 35 Figure 3 17 Amplitude correction...
Страница 213: ...Measurement Examples RIGOL User s Guide for DSA1000 Series 3 37 Figure 3 18 The frequency response of a filter...
Страница 214: ......