Chapter 1 Disassemble and Assemble
RIGOL
DS2000 Service Guide
1-11
Disassemble the Main Board
Figure 1-10 Disassemble the Main Board
Part Explanation:
①
15 screws (M3*6 pan head torx) fixing the main board and front metal panel.
Disassemble Steps:
1.
Remove the 15 screws (
①
) fixing the main board and front metal panel using the
screw driver (T10).
2.
Remove the 3 nuts and washers at the BNC port at the front metal panel using the
BNC socket and the 5mm hexagonal socket.
Tip
Pass the left keyboard cable and the screen line through the corresponding
holes on the front metal panel, fasten the main board and insert each cable
into the corresponding position before assembling the main board.
①
①
①
①
①
①
①
①
①
①
①
①
①
①
①