Example 9: Pass/Fail Test
When running test function the oscilloscope automatically examines the input signal,
compared to the already built waveform mask. If the waveform “touches” the mask,
a “Fail” occurs, otherwise the test passes. When needed, a programmable output
can be used for external automatic control applications, such as the automatic
failure product kick-out in a production line. The output is built in as a standard
feature, and is optically isolated.
Do the steps as following:
1.
Press Utility
→
Pass/Fail.
2.
Press Enable Test and select ON.
3.
Press Mask Setting
→
Load.
4.
Press Load to recall the saved mask or press
X Mask and Y Mask to adjust the
horizontal limit and vertical limit then press Create Mask to create a new mask.
5.
Press Output to select the expected outputting waveforms.
6.
Press Operate to start the test.
Figure 3- 11
Waveform display
www.
GlobalTestSupply
.com
Find Quality Products Online at: