Chapter 3 Constant Current Test
DP1116A Performance Verification Manual
3-10
voltage and current of the scale under test (refer to Table 3-4).
OUTPut:RANGe 16V (command 1)
APPLy 16.8,10 (command 2)
OUTPut:STATe ON (command 3)
2)
The electronic load is still in short-circuit mode. Make sure that the power
supply is in CC mode. Read and record the measurement value (U
3
) of the
multimeter. Calculate the current I
3
=U
3
/R
M
.
3)
Send the MEASure:CURRent? command via the remote interface. Read and
record the returned current (I
4
).
4)
Calculate the programming accuracy: |I
3
-rated output current of the scale
under test|;
Calculate the readback accuracy: |I
4
-rated output current of the scale under
test|.
5.
Press 32V/5A at the front panel and repeat steps 1 to 4 to test the
programming and readback accuracy at 32V/5A scale. The commands in step 3
and step 4 should be replaced by the following commands respectively.
OUTPut:RANGe 32V
APPLy 33.6,0
OUTPut:STATe ON
and
OUTPut:RANGe 32V
APPLy 33.6,5
OUTPut:STATe ON