Performance Verification Manual 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

DP1116A   

Programmable Linear DC Power Supply 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Feb. 2014 

RIGOL Technologies, Inc. 

Содержание DP1116A

Страница 1: ...RIGOL Performance Verification Manual DP1116A Programmable Linear DC Power Supply Feb 2014 RIGOL Technologies Inc...

Страница 2: ......

Страница 3: ...erial RIGOL shall not be liable for losses caused by either incidental or consequential in connection with the furnishing use or performance of this manual as well as any information contained Any par...

Страница 4: ...re any inputs or outputs Connect the Probe Correctly If a probe is used do not connect the ground lead to high voltage since it has the isobaric electric potential as ground Observe All Terminal Ratin...

Страница 5: ...tmosphere Keep Product Surfaces Clean and Dry To avoid the influence of dust and or moisture in air please keep the surface of device clean and dry Electrostatic Prevention Operate in an electrostatic...

Страница 6: ...struments geerdet Schlie enSie die Erdungsklemmen an keine hohe Spannung an Beachten Sie alle Anschl sse Zur Vermeidung von Feuer oder Stromschlag beachten Sie alle Bemerkungen und Markierungen auf de...

Страница 7: ...b jedweder explosiven Atmosph re zu betreiben Ger teoberfl chen sauber und trocken halten Um den Einflu von Staub und Feuchtigkeit aus der Luft auszuschlie en halten Sie bitte die Ger teoberfl chen sa...

Страница 8: ...work normally and is within specifications Main topics of this manual Chapter 1 Test Overview This chapter mainly introduces the test preparations the recommended test devices and the test precaution...

Страница 9: ...mming and Readback Accuracy 2 12 Chapter 3 Constant Current Tests 3 1 Preparations 3 2 CC Load Regulation Rate CC Load Effect 3 4 CC Linear Regulation Rate CC Source Effect 3 6 CC Ripple and Noise Nor...

Страница 10: ......

Страница 11: ...t Overview RIGOL DP1116A Performance Verification Manual 1 1 Chapter 1 Test Overview Topics of this chapter Test Preparations Voltage and Current Values Recommended Test Devices Test Precautions Test...

Страница 12: ...onnecting the power make sure that the setting of the voltage selector at the rear panel of the power supply matches the AC power supply to be connected DP1116A supports the following input voltages 1...

Страница 13: ...scale of the power supply to specified values Table 1 1 lists the rated output values and maximum output values of the voltage and current of the two scales Table 1 1 Voltage and Current Values of the...

Страница 14: ...eadout Resolution 5 bit RIGOL DM3068 Measure the DC voltage RMS Voltmeter Sensitivity 1 mV Bandwidth 10 MHz Agilent 3458A Measure the RMS ripple and noise Electronic Load Voltage Range 60 Vdc Current...

Страница 15: ...k wire inside the instrument chassis can also cause error about 2 mV called Uoffset When testing the load regulation rate voltage peak peak value and transient response time the test devices are conne...

Страница 16: ......

Страница 17: ...cation Manual 2 1 Chapter 2 Constant Voltage Tests Topics of this chapter Environmental Preparations CV Load Regulation Rate CV Load Effect CV Linear Regulation Rate CV Source effect CV Ripple and Noi...

Страница 18: ...rature about 25 make connections according to the figure below using the devices recommended in Recommended Test Devices During the test please set the voltage of the AC power supply according to the...

Страница 19: ...short circuit devices in Figure 2 1 are used to short circuit the S terminal and terminal as well as the S terminal and terminal respectively Before performing the test press Sense to enable the Sense...

Страница 20: ...d to 16V 10A Set the voltage and current of the scale to be tested according to Table 2 1 Press On Off to enable the output of the power supply 4 Turn on the electronic load Adjust the current of the...

Страница 21: ...rmance Verification Manual 2 5 Test Record Form Table 2 1 CV Load Regulation Rate Record Form Scale 16V 10A 32V 5A Specification 0 01 Volt 2 mV 3 6 mV 0 01 Volt 2 mV 5 2 mV DP1116A Setting Volt V 16 3...

Страница 22: ...to 220 V Note the value set in this step should be accordance with the value selected by the voltage selector in the rear panel of DP1116A 3 Turn on DP1116A Press 16V 10A at the front panel to set the...

Страница 23: ...steps 4 to 7 to test the CV linear regulation rate at 32V 5A scale Test Record Form Table 2 2 CV Linear Regulation rate Test Record Form Scale 16V 10A 32V 5A Specification 0 01 Volt 2 mV 3 6 mV 0 01 V...

Страница 24: ...loscope connect the positive terminal of the oscilloscope probe to the positive terminal of the output scale of the power supply connect the ground terminal of the oscilloscope probe to the negative t...

Страница 25: ...pling mode to AC and the input impedance to 1 M turn on the 20 MHz bandwidth limit Enable the peak peak value measurement function of the oscilloscope Make sure that the power supply is in CV mode Rea...

Страница 26: ...f load or vice versa Test Procedures 1 Turn off DP1116A Connect DP1116A AC power supply electronic load and oscilloscope according to Figure 2 1 Make sure that the programmable electronic load is in C...

Страница 27: ...value of t namely X displayed on the screen as shown in the figure below 6 Press 32V 5A at the front panel and repeat steps 1 to 5 to test the transient response time at 32V 5A scale Test Record Form...

Страница 28: ...6A 3 Measure the CV programming accuracy and readback accuracy when the voltage is 0V 1 Turn on DP1116A Send commands via the remote interface to select the scale to be tested set the voltage and curr...

Страница 29: ...and via the remote interface USB LAN or GPIB Read and record the retuned voltage U4 4 Calculate the programming accuracy U3 rated output voltage of the scale under test Calculate the readback accuracy...

Страница 30: ...ing 0 05 Volt 10 mV 26 mV Readback 0 05 Volt 5 mV 21 mV Output voltage is 0 DP1116A Setting Volt V 0 0 Current A 10 5 5 25 Measurement Value Test Result U1 U2 Programming Accuracy U1 0 Readback Accura...

Страница 31: ...Verification Manual 3 1 Chapter 3 Constant Current Tests Topics of this chapter Environmental Preparations CC Load Regulation Rate CC Load Effect CC Linear Regulation Rate CC Source Effect CC Ripple a...

Страница 32: ...se to enable the Sense mode Note that a current sampling resistor RM 0 01 should be connected between the power supply under test and the electronic load serially to convert the current signal under t...

Страница 33: ...Chapter 3 Constant Current Test RIGOL DP1116A Performance Verification Manual 3 3 S S Figure 3 2 4 wire Current Sampling Resistor C C...

Страница 34: ...lected by the voltage selector in the rear panel of DP1116A 3 Turn on DP1116A Press 16V 10A at the front panel to set the scale to be tested to 16V 10A Set the voltage and current of the scale to be t...

Страница 35: ...erification Manual 3 5 Test Record Form Table 3 1 CC Load Regulation Rate Test Record Form Scale 16V 10A 32V 5A Specification 0 005 Current 250 A 750 A 0 005 Current 250 A 500 A DP1116A Setting Volt V...

Страница 36: ...d be accordance with the value selected by the voltage selector in the rear panel of DP1116A 3 Turn on DP1116A Press 16V 10A at the front panel to set the scale to be tested to 16V 10A Set the voltage...

Страница 37: ...to 7 to test the CC linear regulation rate at 32V 5A scale Test Record Form Table 3 2 CC Linear Regulation Rate Test Record Form Scale 16V 10A 32V 5A Specification 0 01 Current 250 A 1 25 mA 0 01 Curr...

Страница 38: ...ep should be accordance with the value selected by the voltage selector in the rear panel of DP1116A 4 Turn on DP1116A Press 16V 10A at the front panel to set the scale to be tested to 16V 10A Set the...

Страница 39: ...ear panel of DP1116A 3 Measure the programming and readback accuracy when the current is 0 A 1 Turn on DP1116A Send commands via the remote interface to select the scale to be tested set the voltage a...

Страница 40: ...he current I3 U3 RM 3 Send the MEASure CURRent command via the remote interface Read and record the returned current I4 4 Calculate the programming accuracy I3 rated output current of the scale under...

Страница 41: ...ramming 0 2 Current 10 mA 20 mA Readback 0 15 Current 5 mA 12 5 mA Output current is 0 DP1116A Setting Volt V 16 8 33 6 Current A 0 0 Measurement Value U1 I1 U1 RM I2 Programming Accuracy I1 0 Readbac...

Страница 42: ......

Страница 43: ...Volt 2 mV 5 2 mV DP1116A Setting Volt V 16 32 Current A 10 5 5 25 Measurement Value U0 U1 U1 U0 CV Linear Regulation Rate Test Record Form Scale 16V 10A 32V 5A Specification 0 01 Volt 2 mV 3 6 mV 0 01...

Страница 44: ...Specification 350 V rms 3 mVpp DP1116A Setting Volt V 16 32 Current A 10 5 5 25 Measurement Value Vpp Vrms Transient Response Time Test Record Form Scale 16V 10A 32V 5A Specification 50 s DP1116A Sett...

Страница 45: ...rement Value Test Result U1 U2 Programming Accuracy U1 0 Readback Accuracy U2 0 Output voltage is the rated output value DP1116A Setting Volt V 16 32 Current A 10 5 5 25 Measurement Value Test Result...

Страница 46: ...6 Current A 10 5 Voltage Selector 100 115 220 230 100 115 220 230 Fuse T4 A T2 5 A T4 A T2 5 A AC Power Supply Setting Undervoltage Vac 90 103 198 207 90 103 198 207 Overvoltage Vac 110 127 242 253 1...

Страница 47: ...0 mA Readback 0 15 Current 5 mA 12 5 mA Output current is 0 DP1116A Setting Volt V 16 8 33 6 Current A 0 0 Measurement Value U1 I1 U1 RM I2 Programming Accuracy I1 0 Readback Accuracy I2 0 Output curr...

Страница 48: ...01 2 mV Current 0 01 250 A Ripple and Noise 20 Hz to 20 MHz Normal Mode Voltage 350 V rms 3 mVpp Normal Mode Current 2 mA rms Annual Accuracy 1 25 5 Output Percentage Offset Programming Voltage 0 05 1...

Страница 49: ...Vac 10 115 Vac 10 220 Vac 10 230 Vac 10 Environment Working Temperature Full Rated Value Output 0 to 40 Under Relatively Higher Temperature the linearity of the output current reduces to 50 at the hi...