47
46
Error Messages and Troubleshooting
Limitations
Meter Malfunction
®
If the
Rightest
Blood Glucose Meter can not be started, please follow the steps below:
1. Open the battery cover, remove the batteries.
2. Wait for 5 minutes and insert the batteries as described “Meter activation and Battery change”
on page 16~17.
®
The
Rightest
Blood Glucose Meter should be working normally after finishing above steps. If not, please
contact your local Bionime customer service.
GM250S
GM250S
Sampling Error
Please don't apply the blood drop to the sample entry of the strip before the
Blood
Glucose Meter displays " ". If you do so, the meter will display " " and " " accompanied by
beeps (if volume is turned on).
Please discard the test strip and repeat the test with a new
Blood Glucose Test Strip.
®
Rightest
®
Rightest
GM250S
GS250S
®
- Store the
Rightest
Blood Glucose Test Strips in the original capped vial at temperatures between 39
℉
to 86
℉
( 4
℃
to 30
℃
) and relative humidity below 90%. Do not freeze.
®
-
Rightest
Blood Glucose Test Strips are designed for use with capillary whole blood samples. Do not use serum
or plasma samples.
- Inaccurate test results may be obtained at high altitude more than about 10000 feet ( 3048 meters ) above sea level.
- Hematocrit below 30% may cause higher results, and hematocrit above 60% may cause lower results.
- Severe dehydration and excessive water loss may cause inaccurately low results.
®
-
Rightest
Blood Glucose Monitoring System has not been validated for use on neonates.
- The glucose test may be interfered under abnormal concentration of
Uric acid
≧
10.0 mg/dL, L-Dopa
≧
3 mg/dL, Dopamine
≧
2 mg/dL, Ascorbic acid (Vitamin C)
≧
5 mg/dL
GS250S
GS250S
GM250S
NOTE
- Acetaminophen, Ibuprofen, Methyldopa, Salicylic Acid, Tetracycline, Tolbutamide, Bilirubin-unconjugated, Cholesterol,
Creatinine, Triglyceride, Maltose, Xylose, Galactose and Lactose ( when occurring in normal blood or normal therapeutic
concentrations ) do not significantly affect results. However, abnormally high concentrations in blood may cause inaccurate results.
- Sodium Fluoride/Potassium Oxalate at low glucose levels may result in inaccurate blood glucose reading. Sodium
Fluoride/Potassium Oxalate should not be used as an anticoagulant with this device.