background image

 

value of active 

(R)

 and reactive 

(X)

 resistance (taking into account the sign) for the series mode 

The third screen displays: 

 

working frequency 

 

|Z| 

value 

 

inductance 

(L)

 and capacitance 

(C)

 value for series mode 

Important! Sometimes inductance or capacitance values are displayed with a minus sign. It would seem to be contrary to 
the laws of physics. But no, in Rig Expert analyzers, the minus sign next to the value indicates how much inductance or 
capacitance you need to add to the circuit in order to compensate for the reactive component of the impedance.

 

On the fourth screen are displayed: 

 

working frequency 

 

magnitude 

 

value of active 

(R)

 and reactive 

(X)

 resistance (taking into account the sign) for parallel mode 

The fifth screen displays: 

 

working frequency 

 

phase value 

 

inductance 

(L)

 and capacitance 

(C)

 value for parallel mode 

Screens switch among themselves by pressing 

  

 keys. 

There are two ways to change the value of the operating frequency: 

 

keystrokes 

 and 

 A single press changes the frequency value minimally. If you press and hold the 

keys, the frequency will change continuously with an accelerated step. 

 

pressing and holding the 

 

 key for 2 seconds. After that, an information window will appear on the analyzer 

screen. Press 

 key again. A window will appear on the analyzer screen with the setting of the operating 

frequency. The 

 

 keys select the cursor position to change units, tens, hundreds, thousands of 

kilohertz, and the 

 

 keys change the value. 

A single press of the 

 key saves the change and returns to 

Single mode

A single press of the 

 key returns to 

Single mode

 without saving changes. 

The help menu is accessible by pressing and holding the 

 key and then selecting the 

Help mode

 with 

the 

 

 keys. 

Multé mode.

 

Multé

 is a new mode, in which you can quickly assess how well your antenna works on different HAM bands. 

The result is displayed as stars. The smaller the 

SWR

 of the antenna in the range, the more stars. 

Table of the dependence of the number of stars on the SWR: 

5

   

–   

SWR from 1.0 to 1.1

 

4

   –   SWR from 1.1 to 1.15

 

3

   –   SWR from 1.15 to 1.3

 

2

   –   SWR from  1.3 to 1.7

 

1

   –   SWR from 1.7 to 3.0

 

The SWR over 3 analyzer in Multé mode ignores. 

If, after scanning, the analyzer did not find a single resonance, then this means that either your antenna has a 
minimum SWR of more than 3, or the antenna or coaxial cable is damaged. 

Содержание Stick 230

Страница 1: ...C Bluetooth Availability Yes Bluetooth ver 4 2 BLE Single mode Class B Battery Charge Time 3 hours RF Power 10 dBm at 50 Ohm load Minimum frequency step 1 kHz Operating temperature 0 40 C 32 104 F Tu...

Страница 2: ...ing measurement mode On the fifth screen when you press the key you will enter the Settings menu The Single mode Single is a measurement of all parameters The mode is started by pressing the key After...

Страница 3: ...will change continuously with an accelerated step pressing and holding the key for 2 seconds After that an information window will appear on the analyzer screen Press key again A window will appear on...

Страница 4: ...he key will bring up the Help menu HAM mode This mode allows you to measure and display the SWR antenna chart on the analyzer screen on one of the fixed HAM radio bands such as 2 2 km 630 m 160 m 80 m...

Страница 5: ...out of range but very close to the beginning or end In such cases the analyzer will determine the resonance beyond the limits of the ranges When you press the or keys the analyzer switches to Free mo...

Страница 6: ...ing the key will cause an accelerated continuous change in frequency long press After press a window will be available with a choice of setting the center frequency exit or calling Help Press briefly...

Страница 7: ...e value Available values 12 5 Ohm 25 Ohm 28 Ohm 37 Ohm 50 Ohm 75 Ohm 100 Ohm 150 Ohm 200 Ohm 300 Ohm 450 Ohm 600 Ohm Default 50 Ohm This setting is needed in cases where for example when measuring a b...

Страница 8: ...The Stick 230 analyzer together with AntScope for Android works with smartphones and tablets running Android version 6 and higher You can download AntScope for Android directly on Google Play https pl...

Страница 9: ...or charging can be connected to any charger for phones tablets with an output voltage of 5 volts and a current of at least 0 5 A The maximum charge current of the analyzer is 1 A The time for a full c...

Страница 10: ...xternal charger The Stick 230 analyzer is protected against battery polarity reversal An incorrectly installed battery will not cause damage to the analyzer and when the charger is connected the analy...

Отзывы: