Press the Green Start key to start the simulation and close the internal valve
within the BP-SIM.
Note when the valve is not closed, the pressure within the cuff will not be
contained and the monitor might go into alarm.
When the simulation is present (ie the internal BP-SIM valve is closed), a “pulse”
symbol will be shown;
5.2.
NIBP Leak Test
In the NIBP leak test, the user is able to test the pneumatic leakage within the
system.
The settings available are:
•
Start Pressure
•
Timer (seconds)
Note: Prior to initiating the test, some NIBP monitors might have to be set
to a service mode to ensure that the internal monitor valves are shut for
incoming pressures. If the monitor’s internal valve is open, the leak test
will not be able to pressurise the system.
Operating Instructions
Rigel BP-SIM