 / P6.5

 

 

 

 

 

 

 

 

8.2002

Trouble Shooting
1.

Motor starter cuts out vacuum pump:

1.1 Check that the incoming voltage and frequency corresponds with the motor data plate.
1.2 Check the connections on the motor terminal block.
1.3 Incorrect setting on the motor starter.
1.4 Motor starter trips too fast.

Solution: Use a motor starter with a time delay trip (version as per IEC 947-4).

1.5 The vacuum pump or the lubricating oil is too cold.
1.6 The viscosity of lubricant is too high.
1.7 Oil mist eliminator elements are blocked or contaminated.
1.8 Back pressure on the exhaust pipework is excessive.
1.8 Continuous operation > 200 mbar (abs.) Solution: Use a bigger motor size.

2.

Insufficient suction capacity:

2.1 Inlet filters or meshes are obscured.
2.2 Suction pipe work is too long or too small.

3.

Vacuum pump does not reach ultimate vacuum:

3.1 Check for leaks on the suction side of the pump or on the system.
3.2 Viscosity of lubricant incorrect.

4.

Vacuum pump operates at an abnormally high temperature:

4.1 Ambient or suction temperature too high.
4.2 Cooling air flow is restricted.
4.3 Problem as per 1.6, 1.7 and 1.8.

5.

Exhausted air contains visible oil mist:

5.1 Oil separator elements are fitted incorrectly.
5.2 Incorrect oil brand is used.
5.3 Problem as per 1.7, 1.8, 4.1 and 4.2.

6.

Unit emits abnormal noise:

Note: A knocking noise from the rotor blades is normal when starting from cold, as long as it disappears within two minutes with increasing
operating temperature.

6.1 The pump cylinder is worn.

Solution: send your complete unit off for repair to the supplier or approved service agent.

6.2 The vacuum regulating valve is noisy.

Solution: replace valve.

6.3 Blades are damaged.
6.4 Problem as per 1.5 and 1.6.

7.

Water in lubricant i.e. Emulsification:

7.1 Pump pulls in water because of the application.

Solution: Fit water separators on to the vacuum side.

7.2 Unit handles more water vapour than the gas ballast is designed for.

Solution: Consult supplier for the provision of an increased gas ballast  capability.

7.3 Pump operates only for a short time and does not reach normal operating temperature.

Solution: Run the pump with closed suction until the oil has been cleaned.

Appendix:

Repair on Site: For all repairs on site an electrician must disconnect the motor so that an accidental start of the unit cannot happen.
All engineers are recommended to consult the original manufacturer or one of the subsidiaries,  agents or service agents. The address of the
nearest repair workshop can be obtained from the manufacturer on application.
After a repair or before re-installation, follow the instructions as
shown under the headings ”Installation and Initial Operation”.
Lifting and Transport: To lift and transport the vacuum pump the
eye bolts on the pump and motor must be used. If an eye bolt is
missing use suitably rated strops. The weight of the pumps is
shown in the accompanying table.
Storage: VC units must be stored in dry ambient conditions with
normal humidity. If a pump needs to be stocked for a period
longer than 3  months we would recommend using an anticorrosion
oil rather than the normal lubricant.
Disposal: The wearing parts (as listed in the spare parts lists)
should be disposed of with due regard to health and safety
regulations.
Spare parts lists:

E 231

VC 100 / VC 150

Noise level (max.)

Weight (max.)

Length

Width

Height

Oil capacity

Содержание VACFOX VC 150

Страница 1: ...er ZMS und Vakuummeter ZVM Verwendung Die Vakuumpumpen VC sind f r den Einsatz im gewerblichen Bereich geeignet d h die Schutzein richtungen entsprechen DIN EN 294 Tabelle 4 f r Personen ab 14 Jahren...

Страница 2: ...enkasten des Motors entf llt bei Ausf hrung mit Stecker Anschluss Die Motordaten sind mit den Daten des vorhandenen Stromnetzes zu vergleichen Stromart Spannung Netzfrequenz zul ssige Stromst rke 4 Mo...

Страница 3: ...Bild Je nach Einsatzh ufigkeit lstand pr fen Erster lwechsel nach 500 Betriebsstunden siehe lablass schraube K Weitere lwechsel nach jeweils 500 2000 Betriebsstunden Bei starkem Staubanfall lwechselin...

Страница 4: ...erventil flattert Abhilfe Ventil ersetzen 6 3 Lamellen sind besch digt 6 4 Fehler wie 1 5 und 1 6 7 Wasser im Schmier l 7 1 Pumpe saugt Wasser an Abhilfe Wasserabscheider vor Pumpe installieren 7 2 Pu...

Страница 5: ...ush coupling Optional extras The following standard optional extras can be supplied if required Vacuum regulating valve ZRV additional non return valve ZRK dust inlet filter ZFP high vacuum suction fi...

Страница 6: ...n be found in the terminal box on the motor unless a special plug connection is fitted Check the electrical data of the motor for compatibility with your available supply voltage frequency permissible...

Страница 7: ...evant retaining clips 2 Lubrication picture Check the oil level regularly depending upon the operating hours First oil change after 500 operating hours see oil drain plug K Further changes every 500 2...

Страница 8: ...d service agent 6 2 The vacuum regulating valve is noisy Solution replace valve 6 3 Blades are damaged 6 4 Problem as per 1 5 and 1 6 7 Water in lubricant i e Emulsification 7 1 Pump pulls in water be...

Отзывы: