Guidelines and manufacturer's declaration - electromagnetic emissions
The e-scope instrument is intended for use in the electromagnetic environment
specified below. The customer or user of the e-scope should ensure that it is used
in such an environment.
Emissions test
Compliance
Electromagnetic environment -
guidance
RF emissions
RF emissions
pursuant to
CISPR 11
Group 1
The e-scope uses RF energy exclusi-
vely for internal functions. Therefore,
its RF emissions are very low and
unlikely to interfere with nearby
electronic devices.
RF emissions
RF emissions
according to
CISPR 11
Class B
The e-scope is intended for use in all
establishments, including residential
areas and those directly connected
to a public supply network that also
supplies buildings used for residen-
tial purposes.
Harmonic
emissions
IEC 61000-3-2
Not applicable
Emissions of
voltage fluctua-
tions, flicker
IEC 61000-3-3
Not applicable