The applicators should never be directed 

to areas sensitive to laser radiation, such 

as the eyes.

• 

It is important that the operator of the 

device verify the electrical installation of 

the device is correct before activating 

the power supply switch.

• 

Laser therapy should not be used over 

areas with tattoos.

• 

Laser therapy should not be used on 

dark pigmented skin.

• 

DO NOT

 operate this device in an 

environment where other devices are 

being used that intentionally radiate 

electromagnetic energy in an unshielded 

manner. Portable and mobile RF 

communication equipment can affect 

Medical Electrical Equipment.

• 

DO NOT

 use sharp objects such as a 

pencil point or ballpoint pen to operate 

the buttons or touch screen as it may 

result in equipment damage not covered 

by the warranty.

• 

DO NOT 

disassemble, modify or 

remodel the device or accessories. This 

could cause the device to malfunction, 

electrical shock, fire or personal injury.

• 

To avoid unauthorized use, it will return 

to the password login after 10 minutes.

INTRODUCTION

Laser therapy is based mostly on 

photochemical and photo-biological effects 

on cells and tissues. Observations have 

shown that if the laser light is supplied 

in the right quantities, you will obtain a 

stimulation of certain cellular functions, 

especially in the presence of cells with 

functional deficiencies. The biological 

action in using Laser therapy produces a 

series of effects on the cells in function 

of a “stimulating” action on mitochondrial 

functions with a higher production of ATP.
Expect the following effects on tissues 

treated with TheraTouch LX2 laser therapy:

• 

Increased local blood circulation

• 

Acceleration of tissue healing 

• 

Stimulation of the fibroblast cells

• 

Increase in collagen deposition

TheraTouch LX2 carries out an important 

therapeutic action for the acceleration 

of the healing process. Furthermore, it 

is indicated for the treatment of painful 

articular, muscle, neurogenic and soft 

tissue syndromes.
The operator, in fact, should be qualified  

to be able to use such equipment, and 

should have passed an adapted training, 

or should operate under the control of a 

medical professional adequately qualified 

to the use of the equipment, in order to 

guarantee safety conditions to the patient.

INDICATIONS

The following indications can be treated 

with TheraToch LX2 laser therapy:

• 

Temporary increase in local blood 

circulation 

• 

Temporary relief of minor muscle and 

joint pain and stiffness

• 

Muscle relaxation

• 

Temporary relief of muscle spasm

• 

Temporary relief of arthritis pain

CONTRAINDICATIONS
• 

Direct radiation in the eye: Class 3B 

Lasers are potentially dangerous for 

the retina. The special safety goggles 

supplied with the machine must be

 

worn by both patient and the operator.

• 

Pregnancy: LLLT should not be 

administered over the low back, 

abdomen or pelvic area during 

pregnancy, 

• 

Neoplasms: you must not use laser 

on a primary lesion or on pain of 

unknown etiology. Laser treatment 

may be granted to relieve pain during 

the terminal stage of disease, it is 

recommended that this be performed 

only with the full consent of the patient.

• 

Thyroid: the laser should never be used 

in this area. 

• 

Over areas of hemorrhage. 

• 

Immuno-suppresive therapy: Any 

patient who has undergone any type of 

drug therapy related to this ailment. 

• 

Treatments over the sympathetic 

ganglia, the vagus nerve and region of 

the heart in patients with heart disease: 

laser therapy can significantly alter 

the neural function, and is therefore 

contraindicated in this regions body 

patients with heart disease.

• 

Atopic dermatitis and eczema in the 

acute phase

4

PRECAUTIONARY INSTRUCTIONS

INTENDED USE

Содержание TheraTouch LX2

Страница 1: ...OPERATION MANUAL TheraTouch LX2...

Страница 2: ...ed by specific symbols Understanding these symbols and their definitions before operating this equipment The definition of these symbols are as follows DANGER Possible safety issues that could be immi...

Страница 3: ...TRUCTIONS 3 INTENDED USE 4 PACKAGE CONTENTS 5 INSTALLATION 6 APPLICATION INFORMATION 6 TREATMENT TIPS 7 OPERATING INSTRUCTIONS 7 DEVICE USER INTERFACE 9 MAINTENANCE 17 SPECIFICATIONS 19 PROTOCOL PARAM...

Страница 4: ...mines in its sole reasonable discretion that the Product does not contain defective workmanship or materials Richmar will inform the Purchaser and return the Product freight billed to the Purchaser NO...

Страница 5: ...the use of or inability to use the product including without limitation damages for loss of goodwill work and productivity computer failure or malfunction or any and all other commercial damages or l...

Страница 6: ...h using any laser therapy device Observe the precautionary and operational stickers placed on the device Use of controls or adjustments or performance of procedures other than those specified herein m...

Страница 7: ...crease in collagen deposition TheraTouch LX2 carries out an important therapeutic action for the acceleration of the healing process Furthermore it is indicated for the treatment of painful articular...

Страница 8: ...energy delivered Frequency Determines the number of laser pulses delivered per second during a pulsed treatment and is measured in Hz Depending on the frequency selected this will determine the amount...

Страница 9: ...ON Do not attempt to use the device if it is not properly grounded Make certain that the device is electrically grounded by connecting it only to a grounded electrical service receptacle conformable w...

Страница 10: ...een taken during the development and in production to obtain the best possible beam characteristics Rough treatment can adversely affect these characteristics and must therefore be avoided To protect...

Страница 11: ...l protocols when device is in standby Favorites Press this button to enter stored favorites when device is in standby Manual Operation Press this button to enter the manual treatment screen when devic...

Страница 12: ...age Display Displays Dosage option 4 measurements available SYSTEM STARTUP STARTUP SCREEN Once power supply is connected turn the On Off Switch on the back of the devices to the On position The Home S...

Страница 13: ...protocol Note To go directly to the Treatment Screen press the desired protocol button TREATMENT INFORMATION SCREEN This screen shows the parameters of treatment and treatment recommendations Touch t...

Страница 14: ...ou do not have the blue interlock connector connected to the back of the device the system will show error 714 Simply plug the connector into the back and proceed with the below directions If the trea...

Страница 15: ...recommend saving under a general treatment description or the patient s chart number Press to save the name chosen under favorites Note Once saved you can view saved Favorites and recall them to be us...

Страница 16: ...gged into each Connection Port Applicator A is on the left of the device and Applicator B is on the right To ensure the correct applicator displayed is also being used for treatment press and the appl...

Страница 17: ...rn orange If you do not have the blue interlock connector connected to the back of the device the system will show error 714 Simply plug the connector into the back and proceed with the below directio...

Страница 18: ...length To erase an incorrect entry and start again press the reset button to clear all the numbers and reenter Once you have the password press enter It will prompt you to re enter the password a seco...

Страница 19: ...ded immediately in the following ways Pressing Pressing Pressing Turn off power switch on the rear of the device FOR FACTORY RESET Press System Settings Press Maintenance Press Factory Reset Input Pas...

Страница 20: ...DANGER Under no circumstances may liquid penetrate the openings on the device i e the connection ports of the cables or interlock connection Therefore do not use cleaning or disinfectant sprays direc...

Страница 21: ...lude 600nm 900nm Measuring range include 10mW 500mW Resolution 5 The brief calibration procedure is as follows 1 Choose a enclosed secure room and put the calibrate device and optical power meter on a...

Страница 22: ...interlock connector 1 Key control Passwords Command of execution of the treatment Button Class of isolation parts applied according to the rule EN 60601 1 Class I B type Dimensions Width Height Depth...

Страница 23: ...d environmental temperature is between 68 77 F 20 25 C Should you receive a cooling off message displayed on the screen please waiting between 2 4 mins before starting the next treatment LASER APPLICA...

Страница 24: ...es all laser diodes or total laser energy Joules J Total Energy Laser LED J cm per diode J diode Pain Inflammation Acute Pulsed 5000Hz 9 0 J 9 4 J 17 3 J cm 1 8 J diode Pain Inflammation Chronic Conti...

Страница 25: ...ustment 22 PROTOCOL PARAMETERS SINGLE APPLICATOR PARAMETERS Protocol Mode Frequency Joules Energy Density J cm Power Density W cm2 per diode Pain Inflammation Acute Pulsed 5000Hz 1 8 J 11 2 J cm 1 1 P...

Страница 26: ...tandard defines the levels of immunity to electromagnetic interferences as well as maximum levels of electromagnetic emissions for medical devices Medical devices manufactured by MANUFACTURER conform...

Страница 27: ...ce in nearby electronic equipment RF emissions CISPR11 Class A The LX2 device is suitable for use in all installations including those in a residential environment and those which are directly connect...

Страница 28: ...typical commercial or hospital environment Surge IEC 61000 4 5 1 kV differential mode 2 kV common mode 1 kV differential mode 2 kV common mode Mains power quality should be that of a typical commercia...

Страница 29: ...mitters as determined by an electromagnetic site survey a should be less than the compliance level in each frequency range b Interference may occur In the vicinity of equipment marked with the followi...

Страница 30: ...ns equipment Rated maximum output Power of transmitter W Separation distance according to frequency of transmitterm 150 kHz to 80 MHz d 1 2 P 80 MHz to 800MHz d 1 2 P 800 MHz to 2 7GHz d 2 3 P 0 01 0...

Страница 31: ...ral symbols Date of manufacture Electrical devices are recyclable material and should not be disposed of with household waste after their useful life Help us to protect the environment and save resour...