DEVICE USER INTERFACE

MANUAL OPERATION CONTINUED

Dosage Selection

• 

Touch the energy density area to highlight it 
orange. Then adjust the density by rotating the 
Central Controller Dial. 

• 

The energy density selected, will change the 
treatment time automatically. You cannot increase 
the treatment time without the device also 
changing the energy density level. 

• 

For a longer treatment time of the selected energy 
density, it is recommended to apply multiple 
applications of the desired dosage, at several 
points. For maximum effectiveness, a minimum of 
20 secs of treatment time is recommended for the 
treatment area.

Treatment Ready

• 

Once the parameters are set, press the start button 
to activate the applicator and the indicator will turn 
orange. 

• 

If you do not have the blue interlock connector 
connected to the back of the device, the system 
will show error 714. Simply plug the connector into 
the back and proceed with the below directions.

• 

If the treatment time equals zero, you cannot 
press the Start button. A treatment time must be 
populated based on the energy density.

Start Treatment

• 

Place firm, comfortable pressure on the treatment 
site with no tilting of applicator head, for maximal 
energy. 

• 

Press the orange button on the applicator to begin 
energy output. The orange button will change 
to blue when energy output has begun. The 
Treatment Time will begin counting down.

• 

During treatment the laser output can be stopped 
by pressing 

 or  .

Dosage Calculator Options

To choose which dosage measurement preferred, 
press the Dosage Calculator Button. There are several 
options to choose from:

• 

Joules only

• 

J/cm

2

 only

• 

J/diode 

• 

All above options in one  
(Joules, J/cm

2

 and J/diode)

Содержание TheraTouch LX2

Страница 1: ...OPERATION MANUAL TheraTouch LX2...

Страница 2: ...ed by specific symbols Understanding these symbols and their definitions before operating this equipment The definition of these symbols are as follows DANGER Possible safety issues that could be immi...

Страница 3: ...TRUCTIONS 3 INTENDED USE 4 PACKAGE CONTENTS 5 INSTALLATION 6 APPLICATION INFORMATION 6 TREATMENT TIPS 7 OPERATING INSTRUCTIONS 7 DEVICE USER INTERFACE 9 MAINTENANCE 17 SPECIFICATIONS 19 PROTOCOL PARAM...

Страница 4: ...mines in its sole reasonable discretion that the Product does not contain defective workmanship or materials Richmar will inform the Purchaser and return the Product freight billed to the Purchaser NO...

Страница 5: ...the use of or inability to use the product including without limitation damages for loss of goodwill work and productivity computer failure or malfunction or any and all other commercial damages or l...

Страница 6: ...h using any laser therapy device Observe the precautionary and operational stickers placed on the device Use of controls or adjustments or performance of procedures other than those specified herein m...

Страница 7: ...crease in collagen deposition TheraTouch LX2 carries out an important therapeutic action for the acceleration of the healing process Furthermore it is indicated for the treatment of painful articular...

Страница 8: ...energy delivered Frequency Determines the number of laser pulses delivered per second during a pulsed treatment and is measured in Hz Depending on the frequency selected this will determine the amount...

Страница 9: ...ON Do not attempt to use the device if it is not properly grounded Make certain that the device is electrically grounded by connecting it only to a grounded electrical service receptacle conformable w...

Страница 10: ...een taken during the development and in production to obtain the best possible beam characteristics Rough treatment can adversely affect these characteristics and must therefore be avoided To protect...

Страница 11: ...l protocols when device is in standby Favorites Press this button to enter stored favorites when device is in standby Manual Operation Press this button to enter the manual treatment screen when devic...

Страница 12: ...age Display Displays Dosage option 4 measurements available SYSTEM STARTUP STARTUP SCREEN Once power supply is connected turn the On Off Switch on the back of the devices to the On position The Home S...

Страница 13: ...protocol Note To go directly to the Treatment Screen press the desired protocol button TREATMENT INFORMATION SCREEN This screen shows the parameters of treatment and treatment recommendations Touch t...

Страница 14: ...ou do not have the blue interlock connector connected to the back of the device the system will show error 714 Simply plug the connector into the back and proceed with the below directions If the trea...

Страница 15: ...recommend saving under a general treatment description or the patient s chart number Press to save the name chosen under favorites Note Once saved you can view saved Favorites and recall them to be us...

Страница 16: ...gged into each Connection Port Applicator A is on the left of the device and Applicator B is on the right To ensure the correct applicator displayed is also being used for treatment press and the appl...

Страница 17: ...rn orange If you do not have the blue interlock connector connected to the back of the device the system will show error 714 Simply plug the connector into the back and proceed with the below directio...

Страница 18: ...length To erase an incorrect entry and start again press the reset button to clear all the numbers and reenter Once you have the password press enter It will prompt you to re enter the password a seco...

Страница 19: ...ded immediately in the following ways Pressing Pressing Pressing Turn off power switch on the rear of the device FOR FACTORY RESET Press System Settings Press Maintenance Press Factory Reset Input Pas...

Страница 20: ...DANGER Under no circumstances may liquid penetrate the openings on the device i e the connection ports of the cables or interlock connection Therefore do not use cleaning or disinfectant sprays direc...

Страница 21: ...lude 600nm 900nm Measuring range include 10mW 500mW Resolution 5 The brief calibration procedure is as follows 1 Choose a enclosed secure room and put the calibrate device and optical power meter on a...

Страница 22: ...interlock connector 1 Key control Passwords Command of execution of the treatment Button Class of isolation parts applied according to the rule EN 60601 1 Class I B type Dimensions Width Height Depth...

Страница 23: ...d environmental temperature is between 68 77 F 20 25 C Should you receive a cooling off message displayed on the screen please waiting between 2 4 mins before starting the next treatment LASER APPLICA...

Страница 24: ...es all laser diodes or total laser energy Joules J Total Energy Laser LED J cm per diode J diode Pain Inflammation Acute Pulsed 5000Hz 9 0 J 9 4 J 17 3 J cm 1 8 J diode Pain Inflammation Chronic Conti...

Страница 25: ...ustment 22 PROTOCOL PARAMETERS SINGLE APPLICATOR PARAMETERS Protocol Mode Frequency Joules Energy Density J cm Power Density W cm2 per diode Pain Inflammation Acute Pulsed 5000Hz 1 8 J 11 2 J cm 1 1 P...

Страница 26: ...tandard defines the levels of immunity to electromagnetic interferences as well as maximum levels of electromagnetic emissions for medical devices Medical devices manufactured by MANUFACTURER conform...

Страница 27: ...ce in nearby electronic equipment RF emissions CISPR11 Class A The LX2 device is suitable for use in all installations including those in a residential environment and those which are directly connect...

Страница 28: ...typical commercial or hospital environment Surge IEC 61000 4 5 1 kV differential mode 2 kV common mode 1 kV differential mode 2 kV common mode Mains power quality should be that of a typical commercia...

Страница 29: ...mitters as determined by an electromagnetic site survey a should be less than the compliance level in each frequency range b Interference may occur In the vicinity of equipment marked with the followi...

Страница 30: ...ns equipment Rated maximum output Power of transmitter W Separation distance according to frequency of transmitterm 150 kHz to 80 MHz d 1 2 P 80 MHz to 800MHz d 1 2 P 800 MHz to 2 7GHz d 2 3 P 0 01 0...

Страница 31: ...ral symbols Date of manufacture Electrical devices are recyclable material and should not be disposed of with household waste after their useful life Help us to protect the environment and save resour...

Страница 32: ...3158 18 10 B 2019 Compass Health Brands Corp Manufactured for Compass Health Brands Corp Toll Free 1 888 549 4945 6753 Engle Road Middleburg Heights OH 44130 richmarweb com...

Отзывы: