TheraTouch® CX2/CX4/EX4

22

PRODUCT DESCRIPTION

The TheraTouch® devices are a family of clinical 

therapy devices which off ers the practitioner a 

wide range of treatment options. These devices 

share an identical control panel equipped with 

a full color touch screen making treatment set-

up easier than ever. A few simple key presses 

are all you need to quick-start a treatment. The 

User Interface intuitively groups and displays 

all the options for a modality setup on the 

large touch screen to ensure that treatment 

parameters can easily be selected and adjusted. 

TheraTouch® EX4

The TheraTouch® EX4 is equipped with two 

or four completely identical electrotherapy 

channels. The electrotherapy channels can 

be used in combination (linked) or totally 

independent. A comprehensive set of current 

waveforms, to include TENS/EMS/IF/Russian/

High Volt, Microcurrent, and RAAS are available, 

targeting both pain management and muscle 

stimulation applications. 
Protocols can run on linked or independent 

channels. With independent channels, two 

or four diff erent protocols can be performed 

simultaneously.

TheraTouch® CX2/CX4

The TheraTouch® CX2/CX4 is a combination 

device, combining the functions of the 

TheraTouch® EX4 and the Ultrasound Therapy 

in a single device. With the TheraTouch® 

CX2/CX4, the simultaneous application of 

ultrasound and electrotherapy (combination 

therapy) is also possible. The remaining 

electrotherapy channels can then be used 

independently.

Electrode Replenishment

For best results and to avoid any adverse 

reactions, it is recommended to use the 
manufacturer supplied electrodes.

 CAUTION

DO NOT

 attach electrodes before device is 

turned on. Device power goes through a self 

test check and could potentially cause shock or 

harm to the patient.

CONNECTION TO AC POWER

Insert the power cable into the socket on the 

back of the device and plug it into a grounded 

wall outlet.

 CAUTION

• 

DO NOT

 place the device in a location where 

the power cord could be tripped over or 

pulled out during treatment.

• 

DO NOT

 attempt to use the device if it is not 

properly grounded. Make certain that the 

device is electrically grounded by connecting 

it only to a grounded electrical service 

receptacle conformable with the applicable 

national and local electrical codes regarding 

medical environments.

• 

Set power switch to “On”. 

 Power LED indicator is lit green indicating 

that the device is connected to the power 

supply.

• 

 The device will initialize and perform a self 

test. This may take a while.

• 

 At the end of the self test, the device enters 

the Home Menu and is ready for use.

DISCONNECTION FROM POWER
• 

When you have fi nished treatments, turn the 

device off  by setting the power switch on 

the back of the device to “Off ”. When the 

green power LED turns off , the device is now 

disconnected from power supply.

HOME SCREEN DISPLAY

Home Screen

When you turn on the unit, you will fi rst enter 
the Home Menu. In the Home Menu, none of the 
channels are selected. The Home Menu provides 
structured access to all therapies available 
within the unit, with appropriate parameter 
defaults. Just select a menu item by touching 
the button to navigate to the next screen. 
Anywhere in the navigation, you can jump back 
to the Home Menu by touching the home 
button. 

TheraTouch® EX4       TheraTouch® CX2/CX4

Содержание DQ8222

Страница 1: ...OPERATION MANUAL TheraTouch CX2 CX4 EX4...

Страница 2: ...ithout prior notice Amendments will however be published in a new edition of this manual 3179 20 01 E 2020 Compass Health Brands Corp All Rights Reserved CONFORMITY TO SAFETY STANDARDS Richmar declare...

Страница 3: ...ulated 2 Pole 9 TENS Biphasic Asymmetrical Symmetrical 10 RAAS Biphasic 11 Russian 12 High Volt 13 Microcurrent 14 NMS 15 Ultrasound Specifications 16 Device Specifications 17 Package Contents 18 Oper...

Страница 4: ...ustment 34 Start Therapy 35 Combination Therapy Set Up Manual Operation 36 Parameters Electrotherapy 36 Treatment Parameter Adjustment 36 Parameters Ultrasound 37 Treatment Time Adjustment 37 Storing...

Страница 5: ...nued supervision of a physician or licensed practitioner These devices have been thoroughly tested and inspected to assure proper performance and operation Specifications put forth in this manual were...

Страница 6: ...mulation device is in use On open wounds or rashes or over swollen red infected or inflamed areas or skin eruptions e g phlebitis thrombophlebitis varicose veins or on top of or in proximity to cancer...

Страница 7: ...the device is on Always turn the power off before removing or changing the electrode location DO NOT leave electrodes attached to the skin after treatment DO NOT lean against or lay on electrodes whi...

Страница 8: ...ltrasonic treatment TheraTouch should be reduced to a comfortable level Moving technique of the applicator should be used when applying therapeutic ultrasound at intensities greater than 0 5 W cm to a...

Страница 9: ...efore each use This device should not be used adjacent to or stacked with other equipment and if adjacent or stacked use is necessary the equipment should be observed to verify normal operation in the...

Страница 10: ...ess heat or fevers Safety has not been established for the use of therapeutic electrical stimulation during pregnancy ULTRASOUND INDICATIONS CONTRAINDICATIONS Indications for Ultrasound Relief of pain...

Страница 11: ...may experience a reaction similar to a heat rash Be sure to inspect the treatment area during and following treatment and discontinue if an adverse reaction does occur Coupling Coupling is described...

Страница 12: ...ncies produces the modulated effect i e Beat H of 4000 and Beat L of 4150 will yield a 150 pps beat frequency Vector Auto Vector Auto is a form of amplitude modulation and is a percentage of the set i...

Страница 13: ...4000 and Beat L of 4150 will yield a 150 pps beat frequency Cycle Time Cycle time refers to the time that the current is On and Off in seconds Example For a cycle time of 10 50 the current will be fl...

Страница 14: ...he frequency to reduce accommodation Example When the pulse frequency is set to 80 Hz and the frequency modulation is set to 40 Hz the final frequency will vary from 80 120 Hz Amplitude Modulation Amp...

Страница 15: ...d period of time and is denoted in Hz or Pulses Per Second pps Burst Duration The time elapsed from the beginning to the end of one burst and is denoted in ms Pattern Frequency The Pattern Frequency r...

Страница 16: ...ty Duty is the percentage of the total treatment time that the current is actually flowing Cycle Time Cycle time refers to the time that the current is On and Off in seconds Example for a Cycle Time o...

Страница 17: ...second period of time and is denoted in Hz or Pulses Per Second pps Polarity This refers to the polarity of the red lead wire connect the lead wire to the active electrode Cycle Time Cycle Time refer...

Страница 18: ...viously been shown to help with reduction in healing time PARAMETERS Frequency In a pulsed current the Frequency refers to the number of pulses that occur in a one second period of time and is denoted...

Страница 19: ...the type of nerve recruited Cycle Time Cycle time refers to the time that the current is on and off in seconds Example for a Cycle Time of 10 50 the current will be flowing for 10 seconds and resting...

Страница 20: ...und output display can be toggled between Watt and Watt cm In pulsed mode the Amplitude during the pulse is displayed The time averaged Amplitude can be obtained by multiplying this value by the Duty...

Страница 21: ...V 50 Hz 60 Hz 1A Power Output DC 15V 4A Max Dimensions 10 x 7 x 5 L x W x H OPERATING ENVIRONMENTAL Temperature 0 C 32 F to 40 C 104 F Relative Humidity 10 90 Atmosphere Pressure 700hPa 1 060hPa TRANS...

Страница 22: ...IC 2 3 x5 MicroBlock Dispersive Electrodes 2 PK 400 852 MIC 1 8 oz Bottle of Therasonic Ultrasound Gel LS5288 1 110 Lead Wires 4 bag WQ8112 1 110 Combination Lead Wire Green WQ8112 SNGL 1 OPTIONAL ACC...

Страница 23: ...nd Indicator Light 13 Ultrasound Head Cradle 14 Ultrasound Head Connection 15 Power Button 16 Carry Handle 17 Grounding Test Post 18 USB Diagnostic Port 19 On Off Switch 20 Power Cord Connection 2 5 3...

Страница 24: ...cm2 Ultrasound Head 14 Ultrasound Indicator Light 15 Ultrasound Head Cradle 16 Ultrasound Head Connection 17 Power Button 18 Carry Handle 19 Grounding Test Post 20 USB Diagnostic Port 21 On Off Switc...

Страница 25: ...s 11 Clinical Protocol Button 12 Favorites Button 13 Power Button 14 Carry Handle 15 Grounding Test Post 16 USB Diagnostic Port 17 On Off Switch 18 Power Cord Connection USB Diagnostic Port is used fo...

Страница 26: ...is recommended to use the manufacturer supplied electrodes CAUTION DO NOT attach electrodes before device is turned on Device power goes through a self test check and could potentially cause shock or...

Страница 27: ...nnels 3 and 4 only available in EX4 CX4 TOUCHSCREEN DISPLAY NAVIGATION BAR Icon Name Function Back Return to previous screen Home Return to Home screen Page Number The number of pages in multi page me...

Страница 28: ...infection electrodes should not be used on broken skin Before Treatment Check the patient for Contraindications as noted under Intended Use section of this manual Test the heat sensitivity of the trea...

Страница 29: ...CC output characteristics may cause unpleasant connection and disconnection reactions if the electrodes are not securely placed or lose contact with the skin Make sure the current amplitude is set to...

Страница 30: ...annels for desired treatment When channel 1 is selected channels 2 3 and 4 are still available for another therapy When channels 1 2 or 3 4 are selected both channels have the same parameters Only the...

Страница 31: ...py cont To pause the treatment touch the pause button in the navigation bar To continue the treatment touch the run button in the navigation bar To stop the treatment touch the STOP button in the navi...

Страница 32: ...hannel 1 is selected channels 2 3 and 4 are still available for another therapy When channels 1 2 or 3 4 are selected both channels have the same parameters Only the intensity can be set differently N...

Страница 33: ...ustment Touch the timer The color will change to orange Adjust the treatment time with the central control dial Start Therapy Once your parameters and timers are set start the therapy by touching the...

Страница 34: ...ical contact You may experience this when the applicator only partially contacts the body When contact restoration is sensed the treatment is resumed at the set amplitude The contact control function...

Страница 35: ...o be adjusted The amplitude can be reduced or the continuous mode can be changed to pulsed mode or vice versa When there are signs that the ultrasound transmission is bad add more contact gel or sprea...

Страница 36: ...l control dial to scroll through the information as there may be several pages of additional information regarding desired clinical protocol Parameters Adjust the parameters by touching the desired pa...

Страница 37: ...n the navigation bar To continue the treatment touch the run button in the navigation bar To stop the treatment touch the STOP button in the navigation bar NOTE To restart after pressing stop you will...

Страница 38: ...nical Protocols Favorites Manual Operation Select Manual Operation button Parameters Adjust the parameters by touching the desired parameter button and change the value with the central control dial N...

Страница 39: ...ent area Then start the therapy by touching the output intensity on the bottom of the screen The color will change to orange Use the central control dial to adjust the output level NOTE The current in...

Страница 40: ...gives access to all functions of the unit Select the Combination button Manual Operation The Combination Menu provides access to the following functions Favorites Manual Operation Select Manual Opera...

Страница 41: ...ng the desired parameter button and change the value with the central control dial Storing Your Favorite Settings When a treatment screen is completely set as required you can save those settings as a...

Страница 42: ...to return to the Home Menu Adjusting Current Amplitude The unity of the displayed current amplitude depends on the previously selected current waveform and can be expressed in mA A or V by pressing t...

Страница 43: ...s Periodically wipe the lead wires clean with a cloth dampened in a mild soap solution and then gently wipe them dry Use of rubbing alcohol on the lead wires will damage the insulation and dramaticall...

Страница 44: ...Restart the device Contact manufacturer tech support 118 External stim board self check has failed 119 Stim board may be defective 120 The storage for favorites is full Delete some of the favorite pro...

Страница 45: ...irections on the electrode packaging for the care of the electrodes The life of an electrode varies depending on skin conditions skin preparation storage and climate Replace electrodes that no longer...

Страница 46: ...Australia Canada and the United States Safety Class According to IEC 60601 1 Class I type BF IEC 60601 2 5 Particular requirements for the safety of ultrasonic therapy equipment IEC 60601 2 10 Partic...

Страница 47: ...TheraTouch unit should ensure that it is operated in such an environment Interference Tests Conformity Electromagnetic Environment Guideline RF emissions according to CISPR 11 Group 1 The TheraTouch...

Страница 48: ...de 2 kV common mode 1 kV differential mode 2 kV common mode Power outlet quality should be that of a typical commercial or hospital environment Voltage dips short interruptions and voltage variations...

Страница 49: ...Hz 3V effective value 150 kHz to 80 MHz 10 V m 80 MHz to 2 7 GHz Portable and mobile radio should not be used any closer to the TheraTouch devices including cables than the recommended separation dist...

Страница 50: ...for operation in an electromagnetic environment where RF disturbances are monitored The customer or user of the TheraTouch Professional Series device can help prevent electromagnetic interference by...

Страница 51: ...mar will replace the product with a new or factory certified refurbished product at Richmar s expense or refund the purchase price to the original purchaser for the price of the defective product minu...

Страница 52: ...3179 20 01 E 2020 Compass Health Brands Corp Manufactured for...