![Richmar APT Performer Скачать руководство пользователя страница 26](http://html1.mh-extra.com/html/richmar/apt-performer/apt-performer_operation-manual_1454539026.webp)
TABLE: TEST SPECIFICATIONS FOR ENCLOSURE PORT IMMUNITY TO RF
WIRELESS COMMUNICATIONS EQUIPMENT
Test
frequency
(MHz)
Band
(MHz) a)
Service a)
Modulation b)
Maximum
power (W)
Distance (m)
Immunity
TEST LEVEL
(V/m)
385
380 to 390
TETRA 400
Pulse
Modulation b):
18Hz
1.8
0.3
27
450
430 to 470
GMRS 460,
FRS 460
FM c) ± 5 Hz
deviation
1 kHz sine
2
0.3
28
710
745
780
704 to 787
LTE Band
13, 17
Pulse
Modulation b):
217 Hz
0.2
0.3
9
810
870
930
800 to 960
GSM
800/900,
TETRA 800,
iDEN 820,
CDMA 850,
LTE Band 5
Pulse
Modulation b):
18 Hz
2
0.3
28
1720
1845
1970
1700 to 1990
GSM 1800;
CDMA 1900;
GSM 1900;
DECT;
LTE Band 1, 3,
4, 25; UMTS
Pulse
Modulation b):
217 Hz
2
0.3
28
2450
2400 to 2570
Bluetooth,
WLAN,
802.11 b/g/n ,
RFID 2450,
LTE Band 7
Pulse
Modulation b):
217 Hz
2
0.3
28
5240
5500
5785
5100 to 5800
WLAN 802.11
a/n
Pulse
Modulation b):
217 Hz
0.2
0.3
9
NOTE:
a) For some services, only the uplink frequencies are included.
b) The carrier shall be modulated using a 50 % duty cycle square wave signal.
c) As an alternative to FM modulation, 50 % pulse modulation at 18 Hz may be used
because while it does not represent actual modulation, it would be worst case.
24
APPENDIX A - EMC TABLES