Installation and Maintenance

 

3

 

NOTE:

 

 Load cell adjustment nut has both right

and left hand threads.

 

9. Lift load cell so that the threaded adapter in

top of load cell lines up with load cell
coupling nut. Turn load cell coupling nut
counterclockwise (looking down from top)
u n t i l   t o p   o f   l o a d   c e l l   c o u p l i n g   n u t   i s
approximately 1/4 inch from jam nut.

10. Wi t h   s c a l e   o n   a   l eve l   s u r fa c e ,   r e m ove

weighing platter from bottom scale assembly.
Place a level across standoffs on levers and
adjust load cell coupling nut until level reads
plumb. Tighten jam nut-to-load cell coupling
nut. Remove level and replace weighing
platter on scale. 

11. Install load cell cable up through hole in

indicator stand/base plate assembly.

 

Do not twist, bend or kink
load cell cable.

 

12. Wrap excess cable around cable hooks on

handle.

 

13. Connect cable from the load cell through the

load cell cable cord grip to connector J1 as
shown in Table 2-1. 

14. If using 6-wire load cell cable (with sense

wires), remove jumpers JP1 and JP2 before
reinstalling connector J1 (see Figure 2-3). For
4-wire installation, leave jumpers JP1 and JP2
on.

15. When connections are complete, reinstall

connector J1 on the board and reinstall
indicator back cover.

16. Install cable tie on wrapped up load cell cable

allowing sufficient cable to be able to tilt
indicator head.

The scale is now ready for final adjustment and
calibration.

 

2.3

Final Adjustment, Calibration and
Troubleshooting

 

Final adjustment, calibration, and troubleshooting of
the IQ plus 390 indicator can be found in the 

 

IQ plus

390-DC Installation Manual

 

.

 

Figure 2-3. IQ plus 390-DC Indicator Board

!

Caution

 

J1 Pin

Function

1

+SIG

2

–SIG

3

+SENSE

4

–SENSE

5

SHIELD

6

+EXC

7

–EXC

For 6-wire load cell connections, remove 
jumpers JP1 and JP2.

 

Table 2-1. J1 Pin Assignments

1

J3

VR2

XT2

VR3

U12

XT1

U2

JP3

U5

U10

JP1

U1

JP2

C27

C26

C28

C29

C30

C31

J1

U14

R1

C4

C3

C7

R6

U7

C19

C18

R7

C16

R8

R9

C20

C21

D2

R29

R26

R25

C45

C46

C47

R23

C48

C49

U13

C42

DR1

R16

L1

L2

1

1

J4

J5

U8

U3

FLASH RAM

U6

U11

C25

VR1

C15

C22

R10

C9

C10

C11

C14

U9

R11

R13

R12

C24

C58

C65

C64

R44

R45

C63

R42

R43

R41

R40

C62

C61

R38

R39

R28

C56

C60

R36

R37

R35

R34

C59

R30

R31

EMI1

F1

J6

C13

C12

C8

R4

C23

C57

C5A

C44

C41

C43

C53

C52

C17

R24

C51

C6

C5

R3

U4

C2

C1

R2

–EXC

+EXC

SHIELD

–SENSE

+SENSE

–SIG

+SIG

7

6

5

4

3

2

1

LOAD CELL
CONNECTOR

R14

R15

DR2

C34

C35

C36

C37

R19

R20

C33

C38

C39

C40

C32

R17

R18

R21

R22

Keypad Connector

Battery

Input

AC Adapter

Input

To Setup

Switch

250 mA

TR5 Fuse

Microcontroller

+

R5

R10A

R10B

Q1

C50

+

C54

+

C54A

+

D1

C58A

R26B

Q2

R26A

C55

C55A

R27

VR4

Отзывы: