Calibration
7
3.0
Calibration
1. Turn the power off.
2. Turn the power back on. During the self-check counting from 9 to 0 press
.
pn
(password)
displays.
3. Press ,
and .
p0chk
is displayed.
4. Press
until
P3 cal
is displayed.
5. Press
to
enter.
6. Press
to
select
CAL
.
7. Press
to enter calibrate.
Nonlin
displays.
3.0.1
Normal Calibration
1. Press
to enter calibrate.
unLd
will display.
2. Remove weight from the scale.
3. When stable indicator is lit, press
. The last calibration weight is displayed.
• If correct, continue by pressing
.
• If not correct, use the
,
and
to change the calibration weight value. When it is
correct, press
.
4. When
LoAd
displays, place the calibration weight on the scale.
5. When stable indicator is lit, press
.
3.0.2
Linearity Calibration
1. Press
to enter calibrate.
2. Press .
Liner
will display.
3. Press
to
enter.
PIN
will display.
M+
GROSS NET
B/N
ESC
TARE
TARE
ZERO
TARE
ZERO
ZERO
ZERO
ZERO
M+
MR
TARE
ZERO
ZERO
ZERO
TARE
ZERO