Configuration
19
CONFIG Menu
Parameter
Choices
Description
GRADS
10000
number
Graduations. Specifies the number of full scale graduations. The value entered must be in
the range 1–999 999 and should be consistent with legal requirements and environmental
limits on system resolution.
To calculate GRADS, use the formula, GRADS =
Capacity
/
Display Divisions
.
Display divisions for primary and secondary units are specified on the FORMAT menu.
ZTRKBN
OFF
0.5D
1D
3D
Zero track band. Automatically zeroes the scale when within the range specified, as long
as the input is within the ZRANGE and scale is at standstill. Selections are ± display
divisions. Maximum legal value varies depending on local regulations.
ZRANGE
1.9%
100%
Zero range. Selects the range within which the scale can be zeroed. The 1.9% selection is
± 1.9% around the calibrated zero point, for a total range of 3.8%. Indicator must be at
standstill to zero the scale. Use 1.9% for legal-for-trade applications.
MOTBAN
1D
2D
3D
5D
10D
20D
OFF
Motion band. Sets the level, in display divisions, at which scale motion is detected. If
motion is not detected for one second or more, the standstill symbol lights. Some
operations, including print and zero, require the scale to be at standstill. Maximum legal
value varies depending on local regulations.
If OFF is selected, ZTRKBN should also be set to OFF.
OVRLOA
FS+2%
FS+1D
FS+9D
FS
Overload. Determines the point at which the display blanks and an out-of-range error
message is displayed. Maximum legal value varies depending on local regulations.
SMPRAT
7.5HZ
15HZ
30HZ
3.75HZ
Sample rate. Selects measurement rate, in samples per second, of the analog-to-digital
converter. Lower sample rate values provide greater signal noise immunity.
DIGFL1
DIGFL2
DIGFL3
8
16
32
64
128
1
2
4
Digital filtering. Selects the digital filtering rate used to reduce the effects of mechanical
vibration from the immediate area of the scale.
Choices indicate the number of A/D conversions that are averaged to obtain the
displayed reading. A higher number gives a more accurate display by minimizing the
effect of a few noisy readings, but slows down the settling rate of the indicator. See
Section 8.6 for more information on digital filtering.
DFSENS
8OUT
16OUT
32OUT
64OUT
128OUT
2OUT
4OUT
Digital filter cutout sensitivity. Specifies the number of consecutive readings that must fall
outside the filter threshold (DFTHRH parameter) before digital filtering is suspended.
DFTHRH
NONE
2DD
5DD
10DD
20DD
50DD
100DD
200DD
250DD
Digital filter cutout threshold. Specifies the filter threshold, in display divisions. When a
specified number of consecutive scale readings (DFSENS parameter) fall outside of this
threshold, digital filtering is suspended. If NONE is selected, the filter is always enabled.