Техническое руководство Rice Lake 150-10-7 доступно для загрузки бесплатно на manualshive.com. Скачайте этот manual, чтобы узнать больше о продукте.
150-10-7
Digital Physician Scale Floor-Level
Software Version 11525
Technical Manual
PN 132098 Rev C
August
1
8, 2022
Страница 1: ...150 10 7 Digital Physician Scale Floor Level Software Version 11525 Technical Manual PN 132098 Rev C August 18 2022...
Страница 2: ...respective companies All information contained within this publication is to the best of our knowledge complete and accurate at the time of publication Rice Lake Weighing Systems reserves the right to...
Страница 3: ...ing or obtained by calling 715 234 9171 and asking for the training department Revision History This section tracks and describes the current and previous manual revisions for awareness of major updat...
Страница 4: ...3 2 Weighing 6 3 3 Hold Release Function 6 3 4 Using the Body Mass Index BMI Function 6 3 4 1 LB Mode 6 3 4 2 KG Mode 7 3 5 Troubleshooting 8 4 0 Configuration 9 4 1 Setup Switch 9 4 2 General Naviga...
Страница 5: ...when the equipment is operated in a commercial environment This equipment generates uses and can radiate radio frequency energy and if not installed and used in accordance with the instruction manual...
Страница 6: ...t the scale while someone is on the scale Do not allow minors children or inexperienced persons to operate this scale Do not use in the presence of flammable materials Do not use this product if any o...
Страница 7: ...the scale back Damage caused by improper packaging is not covered by the warranty 2 2 Setting Up Scale Move the scale into the area where the weighing process will occur Place the scale on a hard lev...
Страница 8: ...e Lake Weighing Systems voids the warranty Figure 2 2 Power Connection Site The battery annunciator on the display turns off when using an AC power connection The brightness of the backlight is reduce...
Страница 9: ...I Pressing the BMI key enables access to the BMI Body Mass Index mode defaults when scale is turned on The patient is gets on the scale weight stabilizes and press the BMI key The display then asks fo...
Страница 10: ...s weight stabilizes The patient s weight and the HOLD and LOCK annunciators remain on the display when the patient is off the scale 3 Press again to return the scale to zero Pressing will not work whi...
Страница 11: ...ight The LOCK annunciator appears on the display 3 Press The BMI and CM annunciators appear on the display and a default height value of 170 0 cm 170 0 is flashing 4 Use and to adjust the height value...
Страница 12: ...scale E4L BAT Battery low but still usable one bar left on indicator display E4U Battery low and unstable no bars left on indicator display E11 CAL Calibration Error recalibrate scale Err 1 Load cell...
Страница 13: ...enus and or parameters Press to enter a displayed menu and or parameter Press or to scroll through values Press to save the displayed selection and move to the next parameter 4 2 2 Enter Numbers Press...
Страница 14: ...located in the back of the scale to enter the setup parameters for the scale Use a small paper clip small screwdriver or other similar object to press the setup switch 4 Once the setup switch is pres...
Страница 15: ...the Up Down arrow keys to increment decrement numbers 7 When done press ENTER key to move to the next parameter LOAD LOAD This is the amount of weight applied during calibration Can also be changed in...
Страница 16: ...2 5 10 20 50 100 Values in lb 1 2 5 10 20 50 100 200 0 2 0 5 1 0 2 0 5 0 10 0 20 0 0 1 The display toggles between a numeric value and ROUND The decimal point location is set to the DISP parameter dis...
Страница 17: ...the following steps 1 Press the ENTER key to change values 2 Use the Up Down arrow keys to change the available parameters 3 When done press ENTER key to move to the next parameter UNITS UNITS Units...
Страница 18: ...f the scale to enter the setup parameters for the scale Use a small paper clip small screwdriver or other similar object to press the setup switch on the back of the indicator under the cover 4 Prog d...
Страница 19: ...the scale until ID flashes 2 Press again 3 Continue to press to scroll through the various menu items 4 Once complete press again then Done displays 5 Press to start the weighing process Figure 4 5 T...
Страница 20: ...the weight was either in motion or out of range If displaying weight and not BMI the scale will send out the following 21 character string xxxxxxxxx SP uu SP mmmmm SP CR LF Example 60 1 KG PATIENT SP...
Страница 21: ...rotocol Commands Name ESC Character ESC Value with Parameters Description Reading R R Tells PC the scale is sending a reading immediately following this is the value that is sent Example ESC R ESC W02...
Страница 22: ...D Z Zero the scale F Show flash values used for the first flash process L USB On Off not available on USB communication Table 5 5 Maintenance Protocol Commands Escape Protocol Example 1 Escape Protoco...
Страница 23: ...the use of the USB driver Driver protections may need to be temporarily disabled on Windows 10 or later computers to allow for the installation of the USB driver 1 Connect the scale s indicator to a P...
Страница 24: ...personnel Go through the following steps for basic maintenance Check the overall appearance of the entire scale for any obvious signs of damage Inspect the condition of the AC power adapter cord for...
Страница 25: ...ack USB Connection Selectable baud rate default 9600 8 bits No parity 1 stop bit No handshaking Environmental Operating Temperature 50 F to 95 F 10 C to 35 C Storage Temperature 32 F to 122 F 0 C to 5...
Страница 26: ...150 10 7 Digital Physician Scale 22 Visit our website www RiceLake com...
Страница 27: ......
Страница 28: ...54868 USA U S 800 472 6703 Canada Mexico 800 321 6703 International 715 234 9171 Europe 31 0 26 472 1319 Rice Lake Weighing Systems Specifications subject to change without notice www ricelake com PN...