Configuration
©
Rice Lake Weighing Systems
●
All Rights Reserved
13
ALEN
Algorithm initial exponent — Maximum
value 10.
Values above 10, will not let you proceed
and will return to the previous value.
Value (8)
The display toggles between a numeric value and
ALEN
;
If you do not want to change this value, press the
BMI
key to move to the
next setting; Example: from ALEN to ATOUT; If you want to change the
value, use the following steps;
1.Press
ENTER
key to change value starting with the right most digit.
2.Use the Up/Down arrow keys to increment/decrement numbers.
3.When done, press
ENTER
key to move to the next parameter (ATOUT).
ATOUT
Algorithm maximal exponent — Maximum
value is 15.
Values above 15, will not let you proceed
and will return to the previous value.
Value (10)
The display toggles between a numeric value and
ATOUT
;
If you do not want to change this value, press the
BMI
key to move to the
next setting; Example: from ATOUT to TOFF; If you want to change the
value, use the following steps;
1.Press
ENTER
key to change value starting with the right most digit.
2.Use the Up/Down arrow keys to increment/decrement numbers.
3.Press
BMI
key to move to the left.
4.Use the Up/Down arrow keys to increment/decrement numbers.
5.When done, press
ENTER
key to move to the next parameter (TOFF).
TOFF
Auto off timer — Measured in minutes;
0 = always on;
Maximum is 9 minutes;
When using an external power supply, this
parameter is irrelevant
5
4
3
2
1
0
9
8
7
6
The display toggles between a numeric value and
TOFF
;
If you do not want to change this value, press the
BMI
key to move to the
next setting; Example: from TOFF to UNITS;
Press the
ENTER
key to move to the next parameter; (UNITS); If you want
to change the value, use the following steps;
1.Press the
ENTER
key to change values.
2.Use the Up/Down arrow keys to change the available parameters.
3.When done, press
ENTER
key to move to the next parameter (UNITS).
UNITS
Units — Selects the unit of measure; It can
be either Kg/Lb, Kg only or Lb only
KG/LB
KG
LB
The display toggles between unit of measurements and
UNIT
;
If you do not want to change this value, press the
BMI
key to move to the
next setting; Example: from UNITS to OP; If you want to change the value,
use the following steps;
1.Press the
ENTER
key to change values.
2.Press the Up/Down arrow keys to change the value.
3.When done, press the
ENTER
key to move to the next parameter (OP).
OP
Binary options:
OP0 — Live weighing options (0=disable,
1=enable)
OP1 — Communication protocol (0=ESC,
1=maintenance)
OP2 — BMI menu (0=disable, 1=enable)
OP3 — RTC power (0=disable, 1=enable)
OP4 - Semi-Auto-Live — (0=disable,
1=enable)
OP5 - Full calculation — (0=spatial, 1=mul
-
tiply by (0=disable, 1=enable
OP6 - Tare - (0=disable, 1=enable
OP7 - Bat type — (0=dry batteries,
1=rechargeable batteries)
OP8 — OIML mode - (0=disable, 1=enable
OP0
OP1
OP2
OP3
The display toggles between a binary option and
POO-0
;
If you do not want to change this value, press the
BMI
key to move to the
next settings If you want to change the value, use the following step
1.Press
ENTER
to change parameters.
2.Use the Up/Down arrow keys to select the display value you want to
change.
3.Press
BMI
key to move the flashing cursor
a. Use the Up/Down arrows to change the value.
b. Press the
BMI
key to move the flashing cursor.
4.Press
ENTER
key to save all of the display parameters.
SAVE
appears
on the display.
5.Press
ENTER
key again and
DONE
appears indicating that you are now
done entering all of the parameters of the scale.
Press the
BMI
key to cycle back to the top level menu choices, ie: PROG/CALIB/DEF/DONE
Parameter
Description
Choices
Steps
Table 4-1. Configuration Mode Menu (Continued)