Dietary/Fitness Technical and Operating Instructions - Troubleshooting & Testing
25
Press the
Kg-Lb
+
REWEIGH
to exit Test mode.
Power On-Off
START
REWEIGH Key
TEST
Version
BAT
A-D
FACTOR
At
Press and
hold
Power Off
Remove Config/
Calibr. Jumper
Figure 9-1. Test Mode Menu Structure
Table 9-2 lists the various display messages when testing the scale.
NOTE:
<-> means the display toggles between the two values.
Table 9-2. Test Mode Menu
Step
Function
Display
Available Parameters
1
Turn the scale on.
While
StArt
is
displayed, press and
hold the
REWEIGH
key.
This enters into
the test mode of
the scale.
StArt
The scale automatically advances to Step
2.
2
Identifies the
software ID
IdEnt <->
11007
Press the
REWEIGH
key to advance to
the next step.
3
Shows the current
weight value
tESt <-> 0.0
To zero the display, press the
Kg-Lb
key.
To advance to the next step, press the
REWEIGH
key.
4
Shows the current
A/D count
A-d <-> XXXX
To advance to the next step, press the
REWEIGH
key.
5
Checks for current
battery level
bAt <-> XXX or nO bAt
If the
nO bAt
is displayed, there are not
batteries in the unit, or the unit is
operating on its external AC adaptor.
To advance to the next step, press the
REWEIGH
key.
6
This shows the
calibration factor.
FACtOr <-> XXXXX
Press the
REWEIGH
key to return to the
top of the test menu.
To exit test mode, press the
REWEIGH
and
Kg-Lb
key simultaneously, or turn off
the indicator.