 Dietary/Fitness Technical and Operating Instructions - Gravity Correction

17

6.0

Gravity Correction

The Rice Lake Platform Fitness Scale features a function that allows correction for gravity differences between 

locations if it is used in an area different from where it was calibrated.
If you plan to use the scale in a location different from that in which it was calibrated, you will need to first 

determine the gravity values, in meters per second

2

 (m/s

2

) for both the calibration location and the location the 

scale will be used (point-of-use). These two values are entered into the scale during calibration and setup. If you 

do not wish to use gravity correction (ie: the scale is going to be used where it was calibrated), then simply enter 

the same value, such as 9.8062 for both the calibration and point of use gravity values. 

NOTE:

 Do not enter 0 for either value. 

Gravity varies from place to place on the earth and is affected most by the location’s latitude and altitude. Gravity 

decreases as you move closer to the equator, and as you go higher in altitude. Several other factors, including 

tidal affects, local anomalies, and buoyancy, also affect gravity’s pull on an object but to a much lesser extent.
One way to determine gravity values is to use an online service such as those provided by the 

National 

Geophysical Data Center

 (www.ngdc.noaa.gov) for the US or Measurement Canada’s web site, (www.ic.gc.ca). 

These sites allow you to enter latitude and longitude information and receive the estimated gravity values.
Another way is by using the following formula:
g-9.780318(1 + 0.0053024sin

2

L - 0.0000058 sin

2

2L) - .000003085H

Where: g = gravity in m/s

2

L = latitude in radians (multiply degrees by 0.0174533 to get radians)
H = altitude in meters
Once you have the gravity values, make sure they are in meters per second (m/s

2

). For instance, if the value you 

have is in milligals (such as those derived from the National Geophysical Data Center web site), you will need to 

divide your value by 100,000 to get m/s

2

. Your values should be between 9.7 and 9.9 (9.8062 is the gravity value 

for 45

º

 N Latitude at sea level).

The calibration value is entered during the calibration before placing weights on the scale. After calibration, go 

into the programming mode and enter the point of use gravity value. Again, if you do not want to use gravity 

correction, enter the same value in both places.

NOTE: 

The gravity correction function has not been evaluated by NTEP or Measurement Canada, therefore it is up to 

the scale dealer to ensure the device is accurate at the intended point of use. 

Содержание 140-10-7 Series

Страница 1: ...118606 Rice Lake Dietary Fitness Scale 140 10 7 Series Software Revision 11387 Technical and Operating Instructions To be the best by every measure...

Страница 2: ......

Страница 3: ...eypad Print 19 8 2 Standard Remote Protocol configuration option 9 set to 1 19 8 3 ESC Protocol configuration option 9 set to 0 20 8 4 Sample and Explanation of ESC Protocol 22 9 0 Troubleshooting Tes...

Страница 4: ...i Dietary Fitness Technical and Operating Instructions Rice Lake continually offers web based video training on a growing selection of product related topics at no cost Visit www ricelake com webinars...

Страница 5: ...lake com health Rice Lake Weighing Systems is an ISO 9001 registered company Figure 1 1 The Rice Lake Dietary Fitness Scale To prevent injury to individuals and damage to your scale please follow thes...

Страница 6: ...eck for scale alignment put the scale on a flat surface and look at the round spirit level on the base of the scale The scale is aligned when the air bubble is immediately in the center of the circle...

Страница 7: ...arged The LED indicator light on the back of the scale housing will illuminate red during the charging period and change over to green when the battery becomes fully charged Load Cell Connections To g...

Страница 8: ...display are very sensitive so only a gentle pushing motion is required to obtain results The scales have the capability of performing different operations beyond just calculating weight The various o...

Страница 9: ...2 Standard Weight Status BMI Weight Status Below 18 5 Underweight 18 5 24 9 Normal 25 0 29 9 Overweight 30 0 and Above Obese The following examples show weight ranges the corresponding BMI ranges and...

Страница 10: ...until TARE disappears from the display The tare weight is also cancelled when the scale is turned off Use the following steps to enter a tare without placing that item on the scale An example of this...

Страница 11: ...libration modes Remove retaining screws x 4 only 2 shown Remove jumper to gain access to configuration and calibration modes Figure 4 1 Gain Access to the Configuration Calibration Enable Jumper on th...

Страница 12: ...S A TImE XX BAUD XX Software ID SET X OP9 X OP10 X GRAV x xxxx Figure 4 2 Configuration Mode Menu Structure 4 3 Numeric Data Entry Use the Kg Lb key to change the numeric data entry while setting up t...

Страница 13: ...hing mode FULL XX XXX Default 600 Lb Note This value is dependent on the model you have Refer to the serial number label on your scale and set appropriately Use the numeric data entry See Section 4 0...

Страница 14: ...press the REWEIGH key 14 Optional features Option 1 allows the selection of unit of measure UOM in calibration and programming OP1 1 Hold the Kg Lb key for browsing through the options Double click on...

Страница 15: ...lick on the Kg Lb key to change the options value or to advance to the next step press the REWEIGH key 0 Disable 1 Enable Option 7 allows you to either enable or disable the baby scale function OP7 0...

Страница 16: ...scale is not in use Options are between one and 20 minutes This is used when the unit is battery operated t OFF 5 Press and hold the Kg Lb key to scroll through the furthest right hand digit on the di...

Страница 17: ...programming mode 1 Access to that jumper is gained by removing the back cover of the indicator Remove the four back cover retaining screws as shown on the left side of Figure 6 on page 7 With the cove...

Страница 18: ...ramming calibration jumper 14 Dietary Fitness Technical and Operating Instructions Figure 4 3 Deep Programming Mode Menu Structure The following table lists the various display messages and sequence w...

Страница 19: ...of negative weight value time X 0 1 sec 20 2 seconds 7 High Absolute Zero HI Abs 2 Absolute high limit for zero in of FS 8 Low Absolute Zero LO Abs 2 Absolute low limit for zero in of FS 9 High Relat...

Страница 20: ...entry See Section 4 0 for more information on Gravity Correction to change the value To advance to the next step press the REWEIGH key 4 Sets the value of the calibration weight you are going to use...

Страница 21: ...to a much lesser extent One way to determine gravity values is to use an online service such as those provided by the National Geophysical Data Center www ngdc noaa gov for the US or Measurement Cana...

Страница 22: ...certification the unit must not be opened Weights and Measures stickers are used to seal the unit and prevent it from being opened Whoever installs this scale is responsible for placing stickers on t...

Страница 23: ...out of range The scale will send out the following 21 character string xxxxxxxxx SP uu SP mmmmm SP CR LF Where xxxxxxxxx is the weight with decimal point and sign if negative uu is the unit lb or kg...

Страница 24: ...e patient s weight ie W0200 5 means 200 5 lb If the scale is overloaded the scale will return the value of 999 9 Units N Nc This indicates in which unit the values have been taken m metric c constitut...

Страница 25: ...Click on FILES and in the dropdown menu select LOAD CONFIGURATION At this time double click on the file Inbar ini 5 Click on STRINGS and the following screen appears Figure 8 2 Strings to be Used in...

Страница 26: ...gure 8 3 HEX Screen Example Examples of what you would see in the ASCII screen are shown in Figure 8 4 Figure 8 4 ASCII Screen Example 8 4 Sample and Explanation of ESC Protocol When the scale measure...

Страница 27: ...the operator will have to ask the scale to send the error data if any exists This is done with the Diagnostics request command and will look like this 4 5 4 JBHOPTF CBUUFSZ 01 3 0 DIRECTION OF COMMUN...

Страница 28: ...A D The load cell is not connected properly Check the cables and mechanical connections If the problem persists replace the set of load cells Err 3 High saturation state high A D See Err 2 Err 6 Unsta...

Страница 29: ...e StArt The scale automatically advances to Step 2 2 Identifies the software ID IdEnt 11007 Press the REWEIGH key to advance to the next step 3 Shows the current weight value tESt 0 0 To zero the disp...

Страница 30: ...h the following steps for basic maintenance 1 Check the overall appearance of the entire scale for any obvious signs of damage abuse etc 2 Inspect the condition of the AC adaptor for cord cracking or...

Страница 31: ...ck Selectable baud rate default 9600 8 bits No parity 1 stop bit No handshaking Environmental Operating Temperature 50 to 104 F 14 to 40 C Storage Temperature 32 to 158 F 0 to 70 C Humidity 85 relativ...

Страница 32: ...les Technical Support 800 472 6703 Canadian and Mexican Customers 800 321 6703 International 715 234 9171 Immediate Emergency Service For immediate assistance call toll free 1 800 472 6703 Canadian an...

Страница 33: ...e nonconformity actually exists and was not caused by accident misuse neglect alteration improper installation improper repair or improper testing RLWS shall be the sole judge of all alleged non confo...

Страница 34: ...PN 118606 11 10...

Отзывы: