User Manual
8
Research Instruments Ltd
IVF Thermometer
Specification & Applicable Part Numbers
Probe Type
Standard microdrop probe
Accuracy
UKAS-certifiedoffset.Measurementatcalibration
points repeatable to ±0.2°C
Sensor
T-type thermocouple
Calibration points
5°C, 37°C, 45°C
Units
Celsius(°C)
Operation time on battery
Approx.500hours,dependingonbatterycapacity
Dimensions
15.5cm x 6.7cm x 4cm
Weight of units
250g
Probe Type
Optional cryo-probe
Tempertaure range
0 to -200°C
Accuracy
±1.5% of reading or 1°C, whichever is greater
Probe length
150mm
Probe diameter
1.5mm
Probe sheath
Stainless steel
Calibration
Notrequired
Probe Type
Optional robust probe
Temperature range
0 to +100°C
Accuracy (calibrated)
±0.2°C(±0.4°F)overtherange-20°Cto+70°C
(-4°Fto+158°F)
0.1%ofreading/±0.2°C(±0.4°F)elsewhere(whichever
isgreater).
Accuracy (uncalibrated)
±0.5°C(±0.9°F)overtherange-20°Cto+70°C
(-4°Fto+158°F)
0.1%ofreading/±0.2°C(±0.4°F)elsewhere(whichever
isgreater).
Calibration
Optional
Part Number
Description
7-90-001
Standard microdrop probe replacement
7-90-011
Optionalcryo-probe
7-90-013
Optionalrobustprobe
7-90-003
Spare probe and holder
7-90-006
Spare probe holder