SDM – Slurry Density Meter
56
www.rhosonics.com
DISTRIBUTORS CALIBRATION (FOR RHOSONICS AND DISTRIBUTORS ONLY)
8.2
mA trimming
For mA trimming a calibrated mA meter is necessary.
The SDM is calibrated at Rhosonics. A step by step instruction is given on the service application
via several pop ups.
8.3
Temperature
Check the temperature reading regularly, and perform a calibration when the reading error
exceeds 5 ºC. It is advised to configure the set temperature with a calibrated sensor, as other
temperature sensors in the same process may lack accuracy or may not be representative for the
temperature at the installation location of the probe.
The SDM doesn’t measure the temperature. The value for the offset is the temperature.
The temperature for the SDM is actually a setting.
Press Temperature button
Press OK when the reference temperature is measured